Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556143278), received between Jul 9, 2025 and Jul 9, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Implant subsidence” (1 report).

1 other recall cover devices cleared under the same FDA submission (K230653, K933958); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Implant subsidence1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2025Implant subsidence; Implant subsidence

    IT WAS REPORTED THAT, AFTER A TKA SURGERY WAS PERFORMED ON (B)(6) 2022, THE TIBIAL COMPONENT HAD SUBSIDED ON THE LATERAL SIDE OF THE TIBIA. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH A LGN POR HA TIB BASE W/O HOLES SZ 6 LT, PRO METAPH TIB STEM 18MM AND LGN CR HIGH FLEX XLPE SZ 5-6 9MM WERE EXCHANGED USING A LEGION REVISION SYSTEM AND UPSIZING THE TIBIAL BASEPLATE, AS THE ORIGINAL WAS UNDERSIZED WHICH CAUSE THE LATERAL SUBSIDENCE. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 22465362 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K230653, K933958. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page