Skip to content

JOURNEY 74027222Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II BCS XLPE ARTICULAR INSERT SIZE 1-2 LEFT 10MM · Class 2 · Implant

All models of JOURNEY (566)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885556177266), received between Mar 3, 2019 and Jan 8, 2024 — classified as 7 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K111711, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Dislodged or Dislocated2
  • Fracture1
  • Insufficient Device Problem Information1
  • Scratched Material1

Patient problems

  • Failure of Implant2
  • Injury2
  • Unspecified Infection2
  • Arthralgia1
  • Bacterial Infection1
  • Fall1
  • Osteolysis1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY WAS PERFORMED, THE PATIENT HAD A FALLEN AND THE CENTRAL STUD HAD BROKEN. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2023, IN WHICH A JRNY II BCS XLPE ART ISRT SZ 1-2 LT 10MM WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Failure of Implant; Fall

    MAUDE report 18465019 ↗
  • InjurySep 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY PERFORMED ON (B)(6) 2017, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO REVISE THE JOURNEY II BCS FEMORAL OXINIUM LEFT SIZE 3 (CASE-(B)(4)), THE JOURNEY TIBIAL BASEPLATE NONPOROUS LT SZ 2 (CASE-(B)(4)), THE JOURNEY II BCS XLPE ARTICULAR INSERT SIZE 1-2 LEFT (CASE-(B)(4)) 10MM AND THE GENESIS II OVAL RESURFACING PATELLAR 29MM (CASE-(B)(4)). THE REASON OF THE REVISION IS UNKNOWN, AS WELL AS THE CURRENT HEALTH STATUS OF THE PATIENT.

    Patient problems: Failure of Implant

    MAUDE report 12543629 ↗
  • InjuryApr 23, 2021Scratched Material; Scratched Material

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2020, THE PATIENT EXPERIENCED AN INFECTION. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. THE JRNY II BCS XLPE ART ISRT SZ 1-2 LT 10MM WAS EXPLANTED; AFTER REMOVING IT, SCRATCHES/PITTING ON THE ARTICULAR SURFACE OF THE MEDIAL AND LATERAL SIDE OF THE IMPLANT WERE NOTICED. OSTEOLYSIS ON THE MEDIAL ANTERIOR PORTION OF THE TIBIA AND ON THE DISTAL MEDIAL AND LATERAL PORTION OF THE FEMUR BONE WAS ALSO NOTICED DURING THE SURGERY, BUT WAS NOT SEEN ON THE XRAY. THE SURGEON MADE A COMMENT ABOUT THE SURFACE OF THE TIBIAL INSERT AND WONDERED WHETHER THAT MAY HAVE BEEN CAUSED BY BONE CEMENT, HOWEVER THERE WERE NO BONE CEMENT PARTICLES PRESENT BY LOOKING AT THE SITE. THE PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Osteolysis; Unspecified Infection

    MAUDE report 11713255 ↗
  • InjuryJul 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED, THE PATIENT SUFFERED FROM INFECTION AND FEVER. THE ADVERSE EVENT WAS TREATED BY REVISING THE JRNY II BCS XLPE ARTICULAR INSERT SZ 1-2 LT 10MM. THE PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Arthralgia; Bacterial Infection

    MAUDE report 10338622 ↗
  • InjurySep 18, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO DISLOCATION. THE INSERT WAS EXCHANGED.

    Patient problems: Injury

    MAUDE report 9083268 ↗
  • InjuryApr 21, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO DISLOCATION. INSERT WAS EXCHANGED. OTHER COMPONENTS WERE RETAINED. NO MORE INFORMATION REPORTED.

    Patient problems: Injury

    MAUDE report 8534332 ↗
  • InjuryMar 3, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    BILATERAL REVISION SURGERY WAS PERFORMED DUE TO INFECTION. INSERT WAS EXCHANGED. THIS IS LEFT SIDE REPORTING.

    Patient problems: Unspecified Infection

    MAUDE report 8384971 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K111711, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-0157-2024 · Sep 27, 2023 · Open, Classified
    JOURNEY II BCS ARTICULAR INSERT, LEFT 10 MM SIZE 5-6, REF 74027262; knee prosthesis
  • Z-0158-2024 · Sep 27, 2023 · Open, Classified
    JOURNEY II BCS CONSTRAINED ARTICULAR INSERT Size 5-6, 10 MM Left, REF 74029262; knee prosthesis
  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1099-2022 · Apr 19, 2022 · Terminated
    LEGION NARROW PS OXIN SZ 6N RT and JRNY II BCS FEMORAL OXIN LT SZ 5

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates