Skip to content

LEGION 71453281Recalls & reported problems

Smith & Nephew, Inc. · LEGION HIGHLY CROSS LINKED POLYETHYLENE DISHED INSERT SIZE 3-4 15MM · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556255674), received between Apr 24, 2019 and Jun 11, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K071071, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Loose or Intermittent Connection1

Patient problems

  • Injury1
  • Joint Dislocation1
  • Laceration(s)1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING A TKA PERFORMED ON (B)(6) 2025, THE PATIENT EXPERIENCED A TIBIAL DISLOCATION WHILE ATTEMPTING TO MOVE HERSELF DOWN THE BED. DR. (B)(6) NOTED THAT THE POSTERIOR CAPSULE APPEARED TO REMAIN INTACT AND ATTRIBUTED THE DISLOCATION TO A LEVERING ACTION DURING SIMULTANEOUS KNEE FLEXION AND INTERNAL ROTATION. THE ADVERSE EVENT WAS ADDRESSED WITH A REVISION SURGERY ON (B)(6) 2025. DURING THIS PROCEDURE, ONE (1) LEGION¿ CR OXIN FEM SZ3 LT AND ONE (1) LGN XLPE DISHED INSERT SZ 3-4 15MM WERE REMOVED. IT WAS DETERMINED THAT A POSTERIOR-STABILISED KNEE IMPLANT WOULD BE MORE APPROPRIATE TO PREVENT RECURRENCE. WHILE REMOVING THE WELL-CEMENTED FEMORAL COMPONENT, A SIGNIFICANT AMOUNT OF BONE REMAINED ATTACHED TO THE IMPLANT. AS A RESULT, A REVISION FEMORAL PROSTHESIS WITH A STEM AND AUGMENTS WAS USED TO COMPENSATE FOR THE BONE LOSS. THE POST ON THE POSTERIOR-STABILISED POLYETHYLENE INSERT SUCCESSFULLY PREVENTED FURTHER DISLOCATION, ALLOWING THE TIBIAL COMPONENT TO BE LEFT IN SITU. PATIENT IS RECOVERING WELL.

    Patient problems: Joint Dislocation; Laceration(s); Unspecified Tissue Injury

    MAUDE report 22184554 ↗
  • InjuryApr 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT RIGHT KNEE REVISION SURGERY WAS PERFORMED DUE TO LOOSE TIBIA.

    Patient problems: Injury

    MAUDE report 8546972 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071071, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1507-2020 · Dec 19, 2019 · Terminated
    LEGION PS High Flex XLPE size 5-6 9mm Articular Insert
  • Z-1904-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 10 MM, REF 71453185 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1905-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 11 MM, REF 71453122 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1907-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 1-2 Size, 9 MM, REF 71453101designed for use in patients in primary and revision surgery, where the anterior and poster
  • Z-1906-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 9 MM, REF 71453121 designed for use in patients in primary and revision surgery, where the anterior and poste

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page