Skip to content

LEGION 71453258Recalls & reported problems

Smith & Nephew, Inc. · LEGION HIGHLY CROSS LINKED POLYETHYLENE POSTERIOR STABILIZED INSERT SIZE 7-8 11MM · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885556255797), received between Feb 11, 2022 and Jan 2, 2026 — classified as 7 injuries.

The most frequently reported specific device problem is “Fracture” (6 reports).

6 other recalls cover devices cleared under the same FDA submission (K071071, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture6
  • Break1
  • Material Fragmentation1

Patient problems

  • Insufficient Information3
  • Failure of Implant2
  • Joint Laxity2
  • No Clinical Signs, Symptoms or Conditions2
  • Pain2
  • Arthralgia1
  • Discomfort1
  • Joint Dislocation1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 2, 2026Fracture; Material Fragmentation

    IT WAS REPORTED THAT AFTER TKA SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE PATIENT´S POST ON THE PS INSERT, BROKE OFF. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION SURGERY ON (B)(6) 2025, IN WHICH THE LGN XLPE PS INSERT SZ 7-8 11MM WAS EXCHANGED. PATIENT'S CURRENT HEALTH CONDITION IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 23950336 ↗
  • InjurySep 4, 2025Fracture; Fracture

    IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2013, THE PATIENT EXPERIENCED THE BREAKAGE OF THE INSERT POST. THIS ADVERSE EVENT WAS TREATED BY REVISION SURGERY ON (B)(6) 2024, IN WHICH THE LGN XLPE PS INSERT SZ 7-8 11MM WAS REVISED TO PSHF POLY. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 22963289 ↗
  • InjuryAug 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY, ONE (1) LGN XLPE PS INSERT SZ 7-8 11MM HAD A BROKEN POST. THIS EVENT WAS TREATED WITH A REVISION SURGERY ON THE (B)(6) 2025 IN WHICH THE INSERT WAS EXCHANGED. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information

    MAUDE report 22840345 ↗
  • InjuryAug 16, 2023Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY PERFORMED ON (B)(6) 2014, THE PATIENT EXPERIENCED PAIN AND SHIFTING WHEN THE KNEE WAS FLEXED. IT WAS DETERMINED THAT THE POST OF THE LGN XLPE PS INSERT SZ 7-8 11MM FRACTURED. IN ORDER TO TREAT THIS, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023, IN WHICH A LGN XLPE PS INSERT SZ 7-8 11MM WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Arthralgia; Failure of Implant; Joint Laxity; Pain

    MAUDE report 17561442 ↗
  • InjuryJul 4, 2022Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA ON (B)(6) 2012, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 TO EXCHANGE THE LGN XLPE PS INSERT SZ 7-8 11MM DUE TO THE POST SHEARING OFF. THE CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14911797 ↗
  • InjuryJun 11, 2022Break; Break

    IT WAS REPORTED THAT, AFTER TKA A A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 DUE TO A LGN XLPE PS INSERT SZ 7-8 11MM BROKE INSIDE THE PATIENT. ALL THE PIECES WERE REMOVED PRIOR TO IMPLANTING A NEW POLY. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14670162 ↗
  • InjuryFeb 11, 2022Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY PERFORMED ON (B)(6) 2013, WHEN ARISING FROM A GROUND POSITION, THE PATIENT'S KNEE WOULD NOT STRAIGHTEN AND IT POPPED. THAT CAUSED PAIN IN THE LATERAL KNEE, IT BECAME SWOLLEN AND WOBBLY. IT WAS DETERMINED THAT THE LEGION HIGHLY CROSS LINKED POLYETHYLENE POSTERIOR STABILIZED INSERT SIZE 7-8 11MM BROKE OFF. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 TO REMOVE THE PIECES. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Discomfort; Failure of Implant; Joint Dislocation; Joint Laxity; Pain; Swelling/ Edema

    MAUDE report 13515632 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K071071, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1507-2020 · Dec 19, 2019 · Terminated
    LEGION PS High Flex XLPE size 5-6 9mm Articular Insert
  • Z-1905-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 11 MM, REF 71453122 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1904-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 10 MM, REF 71453185 designed for use in patients in primary and revision surgery, where the anterior and post
  • Z-1907-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 1-2 Size, 9 MM, REF 71453101designed for use in patients in primary and revision surgery, where the anterior and poster
  • Z-1906-2018 · Mar 30, 2018 · Terminated
    smith&nephew LEGION(R) CR XLPE HIGH LEXION Articular Insert, 5-6 Size, 9 MM, REF 71453121 designed for use in patients in primary and revision surgery, where the anterior and poste