Skip to content

TRIGEN 71674576Recalls & reported problems

Smith & Nephew, Inc. · TRIGEN T-HANDLE · Class 1

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00885556363447), received between Dec 5, 2020 and Feb 8, 2026 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Break” (14 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break14
  • Connection Problem3
  • Crack1
  • Loose or Intermittent Connection1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 8, 2026Break; Mechanical Jam

    IT WAS REPORTED THAT, DURING A TRAUMA TREATMENT, ONE (1) TRIGEN T-HANDLE BROKE AT THE POINT WHERE INSTRUMENTS ARE INSERTED, AND IT DOES NOT MAINTAIN PROPER ATTACHMENT TO INSTRUMENTS THAT ARE INSERTED. THE TREATMENT WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288375 ↗
  • MalfunctionApr 25, 2022Break; Break

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TRIGEN T-HANDLE BROKE. NO PIECES FELL INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT HARMED. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14192481 ↗
  • MalfunctionApr 20, 2022Break; Break

    IT WAS REPORTED THAT, DURING SURGERY, THE CONNECTING PART ON THE TRIGEN T-HANDLE WAS BROKEN. IT IS UNKNOWN HOW THE PROCEDURE WAS RESUMED. NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14161693 ↗
  • MalfunctionSep 15, 2021Break; Break

    IT WAS REPORTED THAT, THE TIP OF A TRIGEN T-HANDLE WAS FOUND TO BE BROKEN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12477969 ↗
  • MalfunctionSep 10, 2021Break; Break

    IT WAS REPORTED THAT, DURING THE SET UP FOR AN INTERNAL FIXATION SURGERY, IT WAS NOTICED THAT THE TIP OF A TRIGEN T-HANDLE WAS FOUND BROKEN. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12455099 ↗
  • MalfunctionSep 10, 2021Break; Break

    IT WAS REPORTED THAT THE TIP OF A TRIGEN T-HANDLE IS BROKEN. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12455066 ↗
  • MalfunctionSep 9, 2021Break; Break

    IT WAS REPORTED THAT ONE TRIGEN T-HANDLE HAS A BROKEN TIP. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12450899 ↗
  • MalfunctionSep 9, 2021Break; Break

    IT WAS REPORTED THAT THE TIP OF A TRIGEN T-HANDLE BROKE AFTER PROCEDURE. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12447411 ↗
  • MalfunctionJul 7, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT A TRIGEN T-HANDLE WILL NOT CONNECT. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12132790 ↗
  • MalfunctionMay 31, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY WHILE THE INSTRUMENTS WERE INSIDE OF THE PATIENT, THE LAG SCREW DRIVER (B)(4) CAPTURE STICK CAUGHT ONTO A GUIDE PIN AND SEEMED TO BE DEFECTIVE; THE PIECE ON THE T-HANDLE (B)(4) WAS BROKEN IN THE CAPTURES OF THE SCREWDRIVER. THERE WERE NO PIECES BROKEN OFF INSIDE OF THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SIGNIFICANT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11909921 ↗
  • MalfunctionMay 28, 2021Break; Break

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE TRIGEN T-HANDLE BROKE OUTSIDE THE PATIENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11904856 ↗
  • MalfunctionMay 18, 2021Break; Break

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE TRIGEN T-HANDLE BROKE OUTSIDE THE PATIENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11842156 ↗
  • MalfunctionMay 13, 2021Break; Break

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE TRIGEN T-HANDLE BROKE OUTSIDE THE PATIENT. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11821004 ↗
  • MalfunctionMay 13, 2021Break; Break

    IT WAS REPORTED THAT, DURING A TRAUMA PROCEDURE A TRIGEN T-HANDLE BROKE. INSTRUMENT OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11816869 ↗
  • MalfunctionMay 13, 2021Break; Break

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE TRIGEN T-HANDLE BROKE OUTSIDE THE PATIENT. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT WAS HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11816846 ↗
  • MalfunctionApr 11, 2021Connection Problem; Connection Problem

    ((B)(4)) IT WAS REPORTED THAT DURING INSPECTION PROCESS, IT WAS FOUND THAT THE THREADS ON THE LAG SCREW DRIVER ARE BENT. THERE WAS NO CASE INVOLVED. ((B)(4)) IT WAS REPORTED THAT DURING INSPECTION PROCESS, IT WAS FOUND THAT THE TRIGEN T-HANDLE DOES NOT CAPTURE AND RELEASE DRIVER PROPERLY. THERE WAS NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11646832 ↗
  • MalfunctionMar 22, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT, A NEW PACKAGED WITH A TRIGEN T-HANDLE WAS OPENED AND IT WAS FOUND THAT THE INSTRUMENT RATTLES, LIKE SOMETHING IS LOSE. IT WAS NOTICES THE 2 BALL BEARINGS MOVING. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11541734 ↗
  • MalfunctionDec 5, 2020Crack; Connection Problem; Crack; Connection Problem

    IT WAS REPORTED THAT THE T HANDLE IS NOT LOCKING AND HAS TWO CRACKS. IT IS UNKNOWN IF THERE WAS ANY DELAY, ANY HARM OR INJURY. IT IS ALSO UNKNOWN HOW THE PROCEDURE WAS FINISHED AND WHEN IT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10951137 ↗

Other INSTRUMENT, BENDING OR CONTOURING devices

All INSTRUMENT, BENDING OR CONTOURING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page