EVOS 71174927Recalls & reported problems

Smith & Nephew, Inc. · T7 DRIVER SHAFT WITH AO QUICK CONNECT · Class 1

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00885556387641), received between Mar 29, 2017 and May 25, 2026 — classified as 3 injuries, 16 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem10
  • Break3
  • Fracture3
  • Appropriate Device Problem Term/Code Not Available1
  • Material Twisted/Bent1
  • Naturally Worn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Device Embedded In Tissue or Plaque2
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 25, 2026Fracture

    IT WAS REPORTED THAT, DURING A PROXIMAL TIBIA ORIF SURGERY, THE TIP OF ONE (1) T7 DRIVER SHAFT W/AO QC BROKE OFF WHILE PUTTING IN A SCREW. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25273369 ↗
  • MalfunctionJul 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, WHILE INSERTING A SCREW INTO A PLATE WITH THE T7 DRIVER SHAFT W/AO QC THE HEAD OF THE DRIVER BROKE. THE PIECES WERE REMOVED AND CONFIRMED VIA C-ARM. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19675040 ↗
  • MalfunctionMay 11, 2022Break; Break

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, THE T7 DRIVER SHAFT W/AO QC BROKE OUTSIDE THE PATIENT. NO PIECES FELL INTO PATIENT. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14362921 ↗
  • InjuryJan 3, 2022Break; Break

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, A T7 DRIVER SHAFT WITH AO QUICK CONNECT SNAPPED WHILE INSIDE THE PATIENT. ALL PIECES WERE RECOVERED. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13142293 ↗
  • MalfunctionFeb 17, 2021Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT, SCREWDRIVERS ARE WORN DOWN SO THEY DON'T FIT IN THE SCREWS PROPERLY. THIS WAS NOTICED DURING INSPECTION. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11341053 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210997 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210962 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210913 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210887 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210867 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210847 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210774 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210738 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THET7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210712 ↗
  • MalfunctionJan 21, 2021Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE T7 DRIVER SHAFT W/AO QC IS NOT CAPTURING THE SCREW AS IT USED TO. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11210686 ↗
  • MalfunctionSep 4, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT BOTH OF THE T8 DRIVER SHAFT W/AO QC AND T7 DRIVER SHAFT W/AO QC SHAFTS SLIPPED INSIDE OF THE SCREW RECESSES WHILE TRYING TO REMOVE HARDWARE. OCCURRED DURING USE-INSIDE THE PATIENT. UNKNOWN IF THERE WAS A SMITH AND NEPHEW BACKUP DEVICE AVAILABLE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10496980 ↗
  • MalfunctionMay 28, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING SURGERY THE END OF THE DRIVER WAS FOUND TO BE ROUNDING OFF. IT HAPPENED OUTSIDE THE PATIENT. NO PATIENT AFFECTATION. NO DELAY REPORTED. THE PROCEDURE COULD BE FINISHED USING AN S&N BACKUP DEVICE.

    Patient problems: No Patient Involvement

    MAUDE report 10099246 ↗
  • InjurySep 9, 2019Break; Break

    IT WAS REPORTED THAT THE DRIVER TIP BROKE OFF DURING SURGERY AND REMAINED INSIDE THE PATIENT. NO DELAY REPORTED. AN S&N BACKUP WAS AVAILABLE.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8976419 ↗
  • InjuryMar 29, 2017Fracture; Fracture

    IT WAS REPORTED THE TIP OF THE DRIVER BROKE DURING A SURGERY AND WAS LEFT IN THE HEAD OF THE SCREW IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 6444098 ↗