Skip to content

JOURNEY 74033661Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II CR INSERT TRIAL LEFT SIZE 5-6 9MM · Class 2

All models of JOURNEY (566)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00885556421857), received between Jun 26, 2019 and Nov 21, 2025 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Fracture6
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY, WHEN THE SURGEON HIT ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M IN, THE FRONT PART CHIPPED OFF. THE PROCEDURE WAS RESUMED WITHOUT ANY DELAY USING A SIMILAR S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23622013 ↗
  • MalfunctionMay 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M BROKE IN HALF. THE SURGEON PULLED ALL THE PIECES FROM THE PATIENT WITH HIS HAND AND ACCOUNTED THEM. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22073900 ↗
  • MalfunctionMay 9, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, WHILE SERVING, A SMALL PIECE OF ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M BROKE OFF. ALL PIECES WERE ACCOUNTED FOR USING A PITUITARY RONGUE. THERE WAS NO SURGICAL DELAY; HOWEVER, IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21994960 ↗
  • MalfunctionJul 19, 2022Fracture; Fracture

    IT WAS REPORTED THAT, THE JRNY II CR ISRT TRL LT SZ 5-6 9M WAS CHIPPED AND UNUSABLE. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15057395 ↗
  • MalfunctionNov 23, 2021Break; Crack; Break; Crack

    IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED AND OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12861443 ↗
  • MalfunctionMar 16, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET UP, BEFORE A TKA THE JRNY II CR ISRT TRL LT SZ 5-6 9M PLASTIC CHIPPED. INCIDENT OCCURRED BEFORE THE PROCEDURE; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11497563 ↗
  • MalfunctionDec 6, 2020Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE JOURNEY II CR INSERT TRIAL BROKE DURING CLEANING PROCESS. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10955750 ↗
  • MalfunctionDec 2, 2020Fracture; Fracture

    IT WAS REPORTED THAT THE IN THE JRNY II CR ISRT TRL LT SZ 5-6 9M IT WAS NOTICED THAT A SMALL PIECE OF MATERIAL HAD COME OFF. NO INFORMATION ABOUT SURGICAL DELAYS, S+N BACKUP DEVICES OR WHEN THIS INCIDENT HAPPENED WAS INITIALLY PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10932131 ↗
  • MalfunctionJan 29, 2020Break; Break

    IT WAS REPORTED THAT DURING SURGERY THE INSERT TRIAL BROKE IN HALF WHILE IT WAS USED FOR TRIALING THE KNEE. NO INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9644076 ↗
  • MalfunctionJul 24, 2019Break; Break

    IT WAS REPORTED THAT THE TRIAL SNAPPED WHILE TRIALING. NO BACKUP DEVICE AVAILABLE. NO MORE INFORMATION PROVIDED YET.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8824272 ↗
  • MalfunctionJun 26, 2019Break; Break

    IT WAS REPORTED THAT DURINGA TKA PROCEDURE THE DEVICE BROKE WHILE SURGEON WAS TRAILING. NO MORE INFORMATION PROVIDED YET.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8737423 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates