JOURNEY 74033661Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY II CR INSERT TRIAL LEFT SIZE 5-6 9MM · Class 2
All models of JOURNEY (566)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00885556421857), received between Jun 26, 2019 and Nov 21, 2025 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Fracture6
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, WHEN THE SURGEON HIT ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M IN, THE FRONT PART CHIPPED OFF. THE PROCEDURE WAS RESUMED WITHOUT ANY DELAY USING A SIMILAR S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23622013 ↗MalfunctionMay 22, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M BROKE IN HALF. THE SURGEON PULLED ALL THE PIECES FROM THE PATIENT WITH HIS HAND AND ACCOUNTED THEM. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22073900 ↗MalfunctionMay 9, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR SURGERY, WHILE SERVING, A SMALL PIECE OF ONE (1) JRNY II CR ISRT TRL LT SZ 5-6 9M BROKE OFF. ALL PIECES WERE ACCOUNTED FOR USING A PITUITARY RONGUE. THERE WAS NO SURGICAL DELAY; HOWEVER, IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21994960 ↗MalfunctionJul 19, 2022Fracture; Fracture
IT WAS REPORTED THAT, THE JRNY II CR ISRT TRL LT SZ 5-6 9M WAS CHIPPED AND UNUSABLE. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15057395 ↗MalfunctionNov 23, 2021Break; Crack; Break; Crack
IT WAS REPORTED THAT, THE LISTED DEVICES WERE CRACKED AND OR BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12861443 ↗MalfunctionMar 16, 2021Break; Break
IT WAS REPORTED THAT, DURING SET UP, BEFORE A TKA THE JRNY II CR ISRT TRL LT SZ 5-6 9M PLASTIC CHIPPED. INCIDENT OCCURRED BEFORE THE PROCEDURE; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11497563 ↗MalfunctionDec 6, 2020Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE JOURNEY II CR INSERT TRIAL BROKE DURING CLEANING PROCESS. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10955750 ↗MalfunctionDec 2, 2020Fracture; Fracture
IT WAS REPORTED THAT THE IN THE JRNY II CR ISRT TRL LT SZ 5-6 9M IT WAS NOTICED THAT A SMALL PIECE OF MATERIAL HAD COME OFF. NO INFORMATION ABOUT SURGICAL DELAYS, S+N BACKUP DEVICES OR WHEN THIS INCIDENT HAPPENED WAS INITIALLY PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10932131 ↗MalfunctionJan 29, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY THE INSERT TRIAL BROKE IN HALF WHILE IT WAS USED FOR TRIALING THE KNEE. NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9644076 ↗MalfunctionJul 24, 2019Break; Break
IT WAS REPORTED THAT THE TRIAL SNAPPED WHILE TRIALING. NO BACKUP DEVICE AVAILABLE. NO MORE INFORMATION PROVIDED YET.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8824272 ↗MalfunctionJun 26, 2019Break; Break
IT WAS REPORTED THAT DURINGA TKA PROCEDURE THE DEVICE BROKE WHILE SURGEON WAS TRAILING. NO MORE INFORMATION PROVIDED YET.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8737423 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page