JOURNEY 74025621Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY II CR INSERT XLPE LEFT SIZE 1-2 9MM · Class 2 · Implant
All models of JOURNEY (566)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00885556422670), received between Jan 15, 2018 and Jun 1, 2026 — classified as 11 injuries.
The most frequently reported specific device problem is “Degraded” (4 reports).
1 other recall cover devices cleared under the same FDA submission (K121443, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded4
- Adverse Event Without Identified Device or Use Problem3
- Device Dislodged or Dislocated1
- Fitting Problem1
- Fracture1
- Loose or Intermittent Connection1
- Material Disintegration1
- Naturally Worn1
Patient problems
- Failure of Implant4
- Pain3
- Insufficient Information2
- Calcium Deposits/Calcification1
- Deformity/ Disfigurement1
- Fall1
- Foreign Body In Patient1
- Implant Pain1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJun 1, 2026Degraded
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2020, ONE (1) JRNY II CR FEM OX NP LT SZ 3 AND ONE (1) JOURNEY TIBIA BASE NP LT SZ 2 EXPERIENCED SURFACE WEAR DUE TO WEAR OF ONE (1) JRNY II CR ISRT XLPE LT SZ 1-2 9MM, RESULTING IN CONTACT BETWEEN COMPONENTS. THIS ADVERSE EVENT WAS TREATED BY REVISION SURGERY ON (B)(6) 2026, IN WHICH THE EXISTING PRODUCTS WILL BE REMOVED AND EXCHANGED WITH COMPETITOR DEVICES. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 25354141 ↗InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2024, THE PATIENT EXPERIENCED AN INFECTION. THIS ADVERSE EVENT WAS TREATED BY REVISION SURGERY ON (B)(6) 2025, IN WHICH ONLY THE JRNY II CR ISRT XLPE LT SZ 1-2 9MM WAS EXCHANGED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Unspecified Infection
MAUDE report 23343934 ↗InjuryJan 10, 2025Degraded; Device Dislodged or Dislocated; Degraded; Device Dislodged or Dislocated
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2023, THE PATIENT EXPERIENCED AN ACCELERATED WEAR THROUGH THE ENTIRE MEDIAL SIDE OF THE BEARING IN LESS THAN ONE YEAR TIME AND WHEN REVISED IT WAS METAL ON METAL. THIS ADVERSE EVENT WAS TREATED BY REVISION SURGERY ON (B)(6) 2024, IN WHICH DEVICE WAS EXPLANTED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Insufficient Information; Joint Dislocation; Joint Laxity; Pain
MAUDE report 21126451 ↗InjuryDec 23, 2023Degraded; Material Disintegration; Degraded; Material Disintegration
IT WAS REPORTED THAT, AFTER A TKA SURGERY USING A JOURNEY SYSTEM, HAD BEEN PERFORMED ON (B)(6) 2019, THE PATIENT EXPERIENCED A FALL AND DEVELOPED KNEE PAIN. ON SUBSEQUENT FOLLOW-UP, X-RAY IMAGE SHOWED CALCIFICATION AND SUSPECTED DEFORMATION OF THE MEDIAL CONDYLE OF THE FEMORAL COMPONENT OR WEAR OF THE INSERT. REVISION SURGERY IS SCHEDULED FOR (B)(6) 2024. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Calcium Deposits/Calcification; Deformity/ Disfigurement; Fall; Implant Pain; Metal Related Pathology; Swelling/ Edema
MAUDE report 18389808 ↗InjuryJul 20, 2023Naturally Worn; Naturally Worn
IT WAS REPORTED THAT, AFTER A TKA WAS PERFORMED ON (B)(6) 2019 WHERE A JRNY II CR ISRT XLPE LT SZ 1-2 9MM, A JOURNEY TIBIA BASE NP LT SZ 2 AND A JRNY II CR FEM OX NP LT SZ 3 WERE IMPLANTED, THE PATIENT EXPERIENCED PAIN. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023. IT WAS FOUND THAT THE INSERT SHOWED WEAR. ALL PRODUCTS WERE EXCHANGED FOR A COMPETITOR'S DEVICES. PATIENT CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant; Pain
MAUDE report 17358440 ↗InjuryMay 4, 2023Fracture; Fracture
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2017, THE PATIENT EXPERIENCED A FRACTURE IN THE IMPLANT. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2023, IN WHICH ONE (1) JRNY II CR ISRT XLPE LT SZ 1-2 9MM WAS REPLACED WITH A S+N BACK-UP DEVICE. HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 16869483 ↗InjuryFeb 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A LEFT KNEE REPLACEMENT SURGERY PERFORMED ON (B)(6) 2017, THE IMPLANTED JRNY II CR ISRT XLPE LT SZ 1-2 9MM WAS NOTED TO BE DEFECTIVE AND WAS EXPLANTED ON (B)(6) 2023. THE PATIENT'S HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 16406339 ↗InjurySep 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A TKA SURGERY WAS PERFORMED ON 2022, PATIENT EXPERIENCED PAIN. ALSO, TIBIAL WAS EXTERNALLY ROTATED. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2022, IN WHICH A JRNY II CR ISRT XLPE LT SZ 1-2 9MM WAS REMOVED. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A COMPETITOR BACK-UP DEVICE INSTEAD. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Pain
MAUDE report 15426592 ↗InjuryJun 10, 2022Degraded; Degraded
IT WAS REPORTED THAT, AFTER A TKA PROCEDURE HAD BEEN PERFORMED ON (B)(6) 2022, THE PATIENT EXPERIENCED SEVERE WEAR OF THE INSERT. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION SURGERY TO EXPLANT THE JRNY II CR ISRT XLPE LT SZ 1-2 9MM, JRNY II CR FEM OX NP LT SZ 4, JOURNEY TIBIA BASE NP LT SZ 2 AND GEN II 7.5MM RESUR PAT 26MM. PATIENT CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant; Foreign Body In Patient
MAUDE report 14655037 ↗InjuryJul 28, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT AFTER THE SURGERY THE PATIENT FELT THE UNCOMFORTABLENESS. X-RAY SHOWED POSTERIOR SUBLUXATION OF THE FEMORAL COMPONENT. SO THE INSERT WAS REVISED.
Patient problems: Injury
MAUDE report 8834669 ↗InjuryJan 15, 2018Fitting Problem; Fitting Problem
THE PATIENT CAME IN EXPERIENCING MID FLEXION INSTABILITY AND IT WAS DETERMINED THAT THE PATIENT COMPROMISED THE PCL. THE PATIENT HAD CR KNEE IN THERE ALREADY AND TO CORRECT THE PROBLEM WE PUT IN A DISHED CR INSERT
Patient problems: Limited Mobility Of The Implanted Joint
MAUDE report 7190819 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121443, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page