ORTHOMATCH 71442165Recalls & reported problems
Smith & Nephew, Inc. · ORTHOMATCH TIBIAL ALIGN M/L TRANS TUBE · Class 2
All models of ORTHOMATCH (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00885556495100), received between Jun 16, 2018 and Jul 2, 2025 — classified as 1 injury, 13 malfunctions.
The most frequently reported specific device problem is “Break” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break14
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Information4
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient2
- Injury1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 2, 2025Break; Break
IT WAS REPORTED THAT, DURING TKA SURGERY, ONE (1) ORTHOMATCH TIB ALIGN M/L TRANS TUBE BROKE WHEN IT WAS BEING POSITIONED IN THE TIBIA. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22391466 ↗MalfunctionFeb 19, 2024Break; Break
IT WAS REPORTED THAT DURING A TKA SURGERY THE ORTHOMATCH TIB ALIGN M/L TRANS TUBE BROKE WHEN ADJUSTING THE ALIGNMENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. ALTHOUGH, THERE WAS A CHANGE IN THE SURGICAL TECHNIQUE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18735735 ↗MalfunctionJan 15, 2024Break; Break
IT WAS REPORTED THAT DURING A ARTHROSCOPY SURGERY THE ORTHOMATCH TIB ALIGN M/L TRANS TUBE BROKE, NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18511211 ↗MalfunctionJul 10, 2022Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY, BEFORE THE ORTHOMATCH TIBIAL ALIGN M/L TRANS TUBE WAS PLACED ON THE PATIENT, IT WAS ASSEMBLED AS REGULAR AND WHEN THE DEVICE WAS SQUEEZED IT JUST CAME OUT IN THE HAND DISASSEMBLING IT AS IT WAS BROKE OFF. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE BY HOLDING IT WITH THE HAND. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14973819 ↗MalfunctionMar 28, 2022Break; Break
IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, AN ORTHOMATCH TIBIAL ALIGN M/L TRANS TUBE WAS OBSERVED TO BE BROKEN. THIS WAS NOTICED OUTSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13927463 ↗MalfunctionFeb 26, 2019Break; Break
IT WAS REPORTED THAT DEVICE BROKE DURING THE PROCEDURE WHILE PERFORMING TIBIAL ALLIGNMENT. BACK UP AVAILABLE. A 30MIN-1 HR DELAY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8370451 ↗MalfunctionJan 4, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY THE TIBIAL ALIGNMENT GUIDE TUBE BROKE WHILST BEING TIGHTENED. DELAY OF LESS THAN 30MIN REPORTED. NO INJURY REPORTED. NO BACK UP REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8218119 ↗MalfunctionNov 12, 2018Break; Break
IT WAS REPORTED THAT DURING PROCEDURE THE ORTOMATCH TIBIAL BOLT WAS BROKEN, NO PROCEDURE, DELAY, INJURY OR BACK UP REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8059527 ↗MalfunctionOct 31, 2018Break; Break
IT WAS REPORTED THAT DURING THE OPERATION OF ANTHEM KNEE JOINT IMPLANTATION, A DISTAL PART OF THE TUBE OF THE TIBIAL GUIDE BROKE.
Patient problems: No Information
MAUDE report 8025857 ↗MalfunctionOct 31, 2018Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A MALFUNCTION OCCURRED. WHEN TIGHTENING THE TIBIAL ALIGNMENT ASSEMBLY DURING TKR, THE PLASTIC ROD BROKE. SURGEON SWITCHED TO GENII METAL TIBIA ALIGNMENT JIG.
Patient problems: No Information; No Known Impact Or Consequence To Patient
MAUDE report 8025315 ↗MalfunctionOct 15, 2018Break; Break
IT WAS REPORTED THAT DURING THE OPERATION OF ANTHEM KNEE JOINT IMPLANTATION, A DISTAL PART OF THE TUBE OF THE TIBIAL GUIDE BROKE. NO DELAY OR INJURY REPORTED. NO BACK UP REPORTED.
Patient problems: No Information
MAUDE report 7965054 ↗MalfunctionSep 26, 2018Break
IT WAS REPORTED THAT DURING THE OPERATION OF ANTHEM KNEE JOINT IMPLANTATION, A DISTAL PART OF THE TUBE OF THE TIBIAL GUIDE BROKE. THERE WERE NO BROKEN PIECES OF THE INSTRUMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7912362 ↗MalfunctionSep 11, 2018Break; Break
IT WAS REPORTED THAT THE INSTRUMENT BROKE DURING THE SURGERY. THERE WAS NO BACKUP DEVICE AVAILABLE. NO PIECES FELL INTO THE PATIENT.
Patient problems: No Information
MAUDE report 7866147 ↗InjuryJun 16, 2018Break; Break
IT WAS REPORTED THAT DURING SURGERY WHEN TIGHTENING THE TIBIAL ALIGNMENT ASSEMBLY, THE PLASTIC ROD BROKE, SUBSEQUENTLY, THE SURGEON HAD TO SWITCHED TO AN INTRA-MODULARY TECHNIQUE TO COMPLETE THE PROCEDURE.
Patient problems: Injury
MAUDE report 7608446 ↗
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page