Skip to content

EVOS 71174986Recalls & reported problems

Smith & Nephew, Inc. · T8 LINEAR DRIVER SHAFT WITH AO QC · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew EVOS T8 driver shaft is a screwdriver tip used to drive screws during orthopedic surgery. Of its 127 reports, 107 are malfunctions and 20 are injuries, with no deaths; many describe the shaft breaking, bending or its tip snapping off during screw placement, with surgeons switching to a back-up device. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
127
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 127 adverse event reports that list this exact device (DI 00885556500590), received between May 1, 2019 and Mar 31, 2026 — classified as 20 injuries, 107 malfunctions.

The most frequently reported specific device problem is “Break” (53 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break53
  • Material Twisted/Bent29
  • Fracture24
  • Material Deformation19
  • Material Fragmentation4
  • Defective Component2
  • Dull, Blunt2
  • Defective Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions83
  • No Consequences Or Impact To Patient17
  • No Known Impact Or Consequence To Patient16
  • No Patient Involvement8
  • No Information5
  • Injury1

Recent adverse event reports

The 25 most recent of 127 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2026Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF WHILE ADVANCING THE SCREW. THE TIP WAS RECOVERED AND DISCARDED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24750674 ↗
  • MalfunctionMar 14, 2026Fracture

    IT WAS REPORTED THAT THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. AS THIS HAPPENED DURING A LAB DEMO, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24600351 ↗
  • MalfunctionAug 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC SNAPPED WHEN IMPLANTING THE 2.7MM LOCKING SCREW. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22862056 ↗
  • MalfunctionAug 18, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY ONE T8 LINEAR DRIVER SHAFT W/ AO QC SNAPPED WHEN IMPLANTING THE 2.7MM LOCKING SCREW, THE PROCEDURE WAS RESUMED WITHOUT ANY SURGICAL DELAY USING A S+N BACK-UP DEVICE AND NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22825110 ↗
  • MalfunctionJan 21, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT, DURING AN ORIF SURGERY, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE WHEN TIGHTENING A SCREW. THE PROCEDURE WAS RESUMED WITHOUT DELAY USING A SMITH AND NEPHEW BACKUP DEVICE, AND NO INJURIES WERE REPORTED AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21203582 ↗
  • MalfunctionDec 31, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE T8 LINEAR DRIVER SHAFT W/ AO QC BROKE WHILE INSERTING THE SCREW. NO PIECES FELL INSIDE THE WOUND, SINCE THE BROKEN PIECE FELL IN THE TABLE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047071 ↗
  • MalfunctionMay 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19233359 ↗
  • MalfunctionApr 8, 2024Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A INTERNAL FIXATION ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC WAS TWISTED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19063709 ↗
  • MalfunctionFeb 20, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION PROCEDURE, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF INSIDE THE HEAD OF A SCREW WHILE FINAL TIGHTENING. THE BROKEN TIP WAS REMOVED OUT OF THE HEAD OF THE SCREW WITH A HEMOSTAT. NO BROKEN PARTS WERE LEFT IN THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18743572 ↗
  • MalfunctionJan 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF A T8 LINEAR DRIVER SHAFT W/ AO QC FRACTURED. IT IS UNKNOWN HOW PROCEDURE WAS COMPLETED, BUT NO DELAY OR PATIENT HARM OCCURRED DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18454090 ↗
  • MalfunctionOct 2, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, A T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF INTO THE SCREW HEAD. IT IS UNKNOWN HOW THIS SURGERY WAS FINISHED OR IF THERE WAS A DELAY. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THE PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17856118 ↗
  • MalfunctionAug 31, 2023Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT THE END OF THE T8 LINEAR DRIVER SHAFT W/ AO QC WAS TWISTED AND BROKE WHILE TIGHTENING THE SCREWS. ALL PIECES RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17666146 ↗
  • MalfunctionAug 29, 2023Break; Break

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE. IT IS UNKNOWN HOW THE TREATMENT WAS COMPLETED AND IF ANY DELAY WAS CAUSED DUE TO THIS INCIDENT. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17642027 ↗
  • MalfunctionAug 15, 2023Break; Break

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THREE (3) T8 LINEAR DRIVER SHAFT W/ AO QC WAS BROKEN. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED OR IF IT CAUSED ANY DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553462 ↗
  • MalfunctionJun 27, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, A T8 LINEAR DRIVER SHAFT W/ AO QC HEAD BROKE OFF WHEN TIGHTENING. THE INSTRUMENT BROKE INSIDE THE PATIENT AND PIECES WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17210936 ↗
  • MalfunctionJun 22, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING SURGERY, WHILE IMPLANTING A 2.7 LOCKING SCREW ON POWER, THE TIP OF A T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. NO PIECES OF THE DRIVER WERE LEFT IN PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17182324 ↗
  • MalfunctionJun 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHILE TIGHTENING A SCREW, THE TIP OF THE T8 LINEAR DRIVER SHAFT W/ AO QC SNAPPED OFF. THE TIP WAS RECOVERED FROM THE PATIENT'S BODY. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE. NO OTHER INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17164091 ↗
  • MalfunctionJun 20, 2023Fracture; Fracture

    IT WAS REPORTED, THAT, DURING AN ANKLE INTERNAL FIXATION. WHEN INSERTING A 2.7 LOCKING SCREW, THE DISTAL TIP OF THE DRIVER BROKE OFF INTO THE TIP OF THE SCREW. THE BROKEN DISTAL TIP WAS RETRIEVED WITH A SET OF PICKUPS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17162285 ↗
  • MalfunctionMay 25, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE TIP OF A T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF IN SCREW WHILE TIGHTENING. A BACK-UP OF THE SAME DEVICE WAS USED AND BROKE AS WELL. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING ONE OF THE BROKEN DEVICES. NO INJURY WAS REPORTED DUE TO THIS ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17007722 ↗
  • MalfunctionApr 13, 2023Break; Break

    IT WAS REPORTED THAT, DURING TRAUMA PLATING SURGERY, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC WAS BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16731741 ↗
  • MalfunctionApr 6, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16690842 ↗
  • MalfunctionMar 3, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE TIP OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16482374 ↗
  • MalfunctionFeb 13, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING TRAUMA SURGERY, THE TIPS OF ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16368308 ↗
  • MalfunctionDec 19, 2022Degraded; Fracture; Material Twisted/Bent; Degraded; Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, ONE (1) T8 LINEAR DRIVER SHAFT W/ AO QC BROKE, STRIPPED AND BENT WHILE PUTTING THE LAST 2.7 SCREW. THE PROCEDURE WAS RESUMED, USING A S+N BACK-UP DEVICE. IT IS UNKNOWN IF A DELAY HAPPENED. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM. NO FURTHER INFORMATION AVAILABLE AT THE MOMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16008645 ↗
  • InjuryNov 3, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, THE TIP OF THE T8 LINEAR DRIVER SHAFT W/ AO QC BROKE OFF. THE BROKEN PIECE WAS RETRIEVED AND NO FRAGMENTS WERE LEFT IN THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15723189 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates