Skip to content

ANTHEM 71422843Recalls & reported problems

Smith & Nephew, Inc. · ANTHEM TIBIAL BASE PLATE SIZE 3 RIGHT · Class 2 · Implant

All models of ANTHEM (25)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556531464), received between Jan 7, 2020 and Jun 10, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

1 other recall cover devices cleared under the same FDA submission (K142807, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Difficult to Insert1
  • Insufficient Device Problem Information1

Patient problems

  • Injury2
  • Anemia1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT DURING A TKA SURGERY, THE ANTHEM TIBIAL BASE PLATE SZ 3 RT COULD NOT LOCK INTO THE ARTICULAR. IT WAS CHANGED THE ARTICULAR TWO TIMES BUT STILL FACED THE SAME ISSUE. THE ANTHEM TIBIAL BASE PLATE SZ 3 RT WAS ALREADY INSERTED INTO PATIENT AND COULD NOT TAKE IT OUT. IT WAS TRIED EVERYTHING BUT STILL COULD NOT LOCK ARTICULAR COMPLETELY SO THIS EVENT COULD CAUSED REVISION IN THE FUTURE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19499217 ↗
  • InjuryFeb 21, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT SUFFERED FROM ANEMIA. THE EVENT WAS TREATED VIA TRANSFUSION AND MEDICATION.

    Patient problems: Anemia

    MAUDE report 9739370 ↗
  • InjuryJan 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ANTHEM CR CLINICAL STUDY 17-4065-07, SUBJECT: 40-022, AE#02. IT WAS REPORTED THAT PATIENT UNDERWENT KNEE MANIPULATION TO RESOLVE A ROM LIMITATION ISSUE.

    Patient problems: Injury

    MAUDE report 9557689 ↗
  • InjuryJan 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(6) CLINICAL STUDY (B)(6). SUBJECT: (B)(6). AE#02: IT WAS REPORTED THAT PATIENT UNDERWENT KNEE MANIPULATION TO RESOLVE A ROM LIMITATION ISSUE.

    Patient problems: Injury

    MAUDE report 9557482 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142807, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page