Skip to content

Legion 71442013Recalls & reported problems

Smith & Nephew, Inc. · LEGION TIBIAL CONE IMPACTOR HANDLE 4MM OFFSET SLEEVE · Class 2

All models of LEGION (446)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556537176), received between Oct 21, 2025 and Oct 22, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K162775); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent2
  • Separation Failure2
  • Difficult to Insert1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2025Difficult to Insert; Separation Failure; Material Twisted/Bent; Difficult to Insert; Separation Failure; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING A LEGION REVISION SURGERY IN THE LEFT KNEE, IT WAS NOTICED THAT ONE (1) LGN TIB CONE IMPCTR HNDL 4MM OFFST SLEEV WAS NOT INSERTING IN ONE (1) LGN CONE IMPACTOR HANDLE. THE SLEEVE WAS NOT DEPRESSING, NOR ENGAGING IN THE HANDLE, IT WOULDN'T FULLY SEAT. IT WAS UNCLEAR WHETHER THIS WAS DUE A FAULT WITH THE SLEEVE OR THE HANDLE, HOWEVER AFTER THE CASE IT WAS NOTICED THE SLEEVE SEEMED BENT. DURING THE CASE THE SURGEON HAD TO HIT THE SLEEVE TO GET IT TO ENGAGE IN THE BROACH HANDLE, THIS ITSELF REQUIRED A LOT OF FORCE; THE SURGEON REPORTED HE HIT HIS FINGERS A NUMBER OF TIMES WHICH WAS PAINFUL. AFTER HITTING THE SLEEVES A NUMBER TIMES, THE SURGEON MANAGED TO SEAT THE SLEEVE IN THE CONE BROACH HANDLE. HE WAS THEN ABLE TO BROACH TO THE REQUIRED CONE SIZE. DUE TO HOWEVER IMPACTING THE SLEEVE INTO THE BROACH HANDLE SO HARD IT BECAME WELDED IN THE BROACH HANDLE WHICH RESULTED IN THE 4MM SLEEVE GETTING STUCK. LATER ON IN THE SURGERY, WHEN PREPARING THE FEMUR, THE SURGEON NEEDED FEMORAL CONES FOR METAPHYSEAL FIXATION. AS NO OFFSET CAN BE USED WITH CONES, THE 4MM OFFSET BEING STUCK IN THE BROACH HANDLE WAS A SERIOUS PROBLEM. THE SURGEON AGAIN MANAGED TO HAMMER THE SLEE

    Patient problems: Pain

    MAUDE report 23360818 ↗
  • MalfunctionOct 21, 2025Positioning Failure; Separation Failure; Material Twisted/Bent; Positioning Failure; Separation Failure; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING A LEGION REVISION SURGERY IN THE LEFT KNEE, IT WAS NOTICED THAT ONE (1) LGN TIB CONE IMPCTR HNDL 4MM OFFST SLEEV WAS NOT INSERTING IN ONE (1) LGN CONE IMPACTOR HANDLE. THE SLEEVE WAS NOT DEPRESSING, NOR ENGAGING IN THE HANDLE, IT WOULDN'T FULLY SEAT. IT HAD TO HIT THE SLEEVE TO GET IT TO ENGAGE IN THE BROACH HANDLE. AFTER HITTING THE SLEEVES A NUMBER OF TIMES, THE SURGEON MANAGED TO SEAT THE SLEEVE IN THE CONE BROACH HANDLE. DUE TO HOWEVER IMPACTING THE SLEEVE INTO THE BROACH HANDLE SO HARD IT BECAME WELDED IN THE BROACH HANDLE WHICH RESULTED IN THE 4MM SLEEVE GETTING STUCK. LATER ON IN THE SURGERY, WHEN PREPARING THE FEMUR, IT WAS NEEDED FEMORAL CONES FOR METAPHYSEAL FIXATION. AS NO OFFSET CAN BE USED WITH CONES, THE 4MM OFFSET BEING STUCK IN THE BROACH HANDLE WAS A SERIOUS PROBLEM. THE SURGEON AGAIN MANAGED TO HAMMER THE SLEEVE OUT. DUE TO THE BROACH HANDLE BEING FAULTY AND ONE (1) LGN CONE IMPCTR HNDL NEUTRAL SLEEVE ALSO NOT DEPRESSING INTO THE HANDLE, THE SURGEON HAD TO BROACH AND INSERT TRIAL AND DEFINITIVE IMPLANT USING THE BROACH WITHOUT A SLEEVE. THE PROCEDURE WAS COMPLETED, AFTER A DELAY OF ONE HOUR, USING THE SAME DEVICES. THIS EVENT INCREASED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349548 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162775. One submission often covers many models, so check the product and lot details in each record.

  • Z-2625-2017 · May 30, 2017 · Terminated
    LEGION(R) TIB CONE IMPACTOR HEADS: a) REF 71441972, ID 24MM b) REF 71441971, ID 22MM c) REF 71441973, ID 26MM d) REF 71441974, ID 28MM e) REF 71441975, ID 30MM
  • Z-2624-2017 · May 30, 2017 · Terminated
    LEGION(R) FEM CONE IMPACTOR HEADS: a) REF 71441976, ID 18MM b) REF 71441977, ID 20MM c) REF 71441978, ID 22MM

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page