ZUK 00584102508Recalls & reported problems
Smith & Nephew, Inc. · Articular Surface Provisional Size 5 8mm · Class 1
All models of ZUK (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew articular surface provisional sizing instrument is used to trial implant sizes during knee surgery. Most of the 32 reports describe it breaking, cracking or chipping, sometimes in half, and in a few cases leaving a piece or fragment in the patient; several were found damaged only during inspection, with no patient involved. Reports mention 6 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 32
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00885556619568), received between Aug 10, 2018 and Jul 21, 2021 — classified as 6 injuries, 26 malfunctions.
The most frequently reported specific device problem is “Break” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break25
- Crack2
- Fracture2
- Material Deformation2
- Material Fragmentation1
- Material Split, Cut or Torn1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Consequences Or Impact To Patient7
- No Known Impact Or Consequence To Patient6
- Foreign Body In Patient2
- No Information1
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2021Crack; Crack
IT WAS REPORTED THAT AN ARTICULAR SURFACE PROVISIONAL SZ 5 8MM WAS FOUND CRACKED. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12201309 ↗MalfunctionJun 22, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY WHILE THE INSTRUMENTS WAS OUTSIDE OF THE PATIENT, THE ARTICULAR SURFACE PROVISIONAL SZ 5 8MM WAS BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12046451 ↗MalfunctionJun 12, 2021Break; Break
IT WAS REPORTED THAT, THE ARTICULAR SURFACE PROVISIONAL SZ 5 8MM BROKE IN HALF. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11988798 ↗MalfunctionMay 7, 2021Fracture; Fracture
IT WAS REPORTED THAT, DURING INSPECTION, AN ARTICULAR SURFACE PROVISIONAL SZ 5 8 MM DEVICE WAS FOUND BROKEN IN HALF. NO CASE INVOLVED; THEREFORE, NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11794660 ↗MalfunctionMay 4, 2021Break; Break
IT WAS REPORTED THAT THE ARTICULAR SURFACE PROVISIONAL SZ 5 8 MM WAS NOTICED BROKEN. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS A PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11770716 ↗InjuryApr 23, 2021Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE BOTTOM OF AN ARTICULAR SURFACE PROVISIONAL SZ 8MM WAS CHIPPED DURING TRIALING, ALL PIECES WERE RETRIEVED. THERE WAS NO SIGNIFICANT DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11720898 ↗MalfunctionApr 20, 2021Break; Break
IT WAS REPORTED THAT, THE ARTICULAR SURFACE PROVISIONAL SZ 5 8MM BROKE. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS A PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11693225 ↗MalfunctionApr 15, 2021Break; Break
IT WAS REPORTED THAT, DURING EFIP INSPECTION, IT WAS NOTICED THAT THE TIBIA SPACER WAS CHIPPED. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11671415 ↗MalfunctionFeb 15, 2021Naturally Worn; Material Split, Cut or Torn; Naturally Worn; Material Split, Cut or Torn
IT WAS REPORTED THAT, DURING INSPECTION THE SZ 5 8MM ZUK NSERT TRL PRESENTED STRONG WEAR/DAMAGE ON THE EXTRUDED TRIAL. NO CASE REPORTED THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. HOWEVER, DURING THE INSPECTION OF THE DEVICE IT WAS CONFIRMED THAT THE DEVICE IS HEAVILY GOUGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11326574 ↗MalfunctionFeb 14, 2021Break; Break
IT WAS REPORTED THAT, AN ARTICULAR SURFACE PROVISIONAL SZ 5 8MM IS CHIPPED. INCIDENT OCCURRED DURING INSPECTION. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11326215 ↗InjuryFeb 12, 2021Break; Break
IT WAS REPORTED THAT, DURING TKA SURGERY THE ZUK SZ. 5 8MM ZUK POLY TRIAL WAS PLACED IN THE PATIENT FOR SIZING AND BROKE IN HALF INSIDE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 11325744 ↗InjuryFeb 4, 2021Break; Break
IT WAS REPORTED THAT, DURING A UKR PROCEDURE, ARTICULAR SURFACE PROVISIONAL SZ 5 8MM DEVICE BROKE IN HALF LENGTH WISE WHEN THE SURGEON WAS INSERTING. THE DEVICE BROKE DURING SURGERY, WHILE INSIDE THE PATIENT. THERE WERE 2 PIECES, NOTHING WAS LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11283557 ↗MalfunctionJan 14, 2021Break; Break
IT WAS REPORTED THAT DURING AN INSPECTION IT WAS FOUND THAT THE ZUK INSERT TRIAL WAS DAMAGED/BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11174730 ↗MalfunctionDec 30, 2020Break; Crack; Break; Crack
IT WAS REPORTED THAT, DURING SURGERY, ARTICULAR SURFACE PROVISIONAL SZ 5 8MM CRACKED IN HALF WHILE INSIDE THE PATIENT. ALL PIECES WERE REMOVED FROM THE PATIENT. A BACK-UP DEVICE WAS AVAILABLE AND NO DELAY OR INJURY OCCURRED DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11097941 ↗MalfunctionDec 29, 2020Break; Break
IT WAS REPORTED THAT WHILE SETTING UP FOR THE SURGERY THE ZUK SZ 5 8MM TRIAL WAS NOTICED TO BE BROKEN. A BACK UP S&N ITEM WAS AVAILABLE, THERE WAS NO DELAY, THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11089262 ↗InjuryDec 9, 2020Break; Break
IT WAS REPORTED THAT, DURING SURGERY, THE TRIAL 8MM INSERT BROKE DURING CEMENTING FOR A UNI KNEE, JZ CONSTRUCT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SMITH+NEPHEW BACK UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10980094 ↗InjuryNov 11, 2020Break; Break
IT WAS REPORTED THAT THE POLY TRIAL BROKE WHILE TRIALING. THE INCIDENT OCCURRED DURING USE, WHILE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE. NO SURGICAL DELAY OR INJURY TO THE PATIENT WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Foreign Body In Patient
MAUDE report 10824994 ↗MalfunctionNov 11, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT TRIAL WAS NOTICED TO BE CHEWED UP. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Patient Involvement
MAUDE report 10823554 ↗InjuryOct 19, 2020Break; Break
IT WAS REPORTED THAT DURING A ZUK CASE, THE ARTICULAR SURFACE PROVISIONAL SZ 5 8MM WAS BREAKING . IT OCCURRED AFTER THE FEMORAL AND TIBIAL IMPLANTS WERE CEMENTED IN. THE SURGEON TRIED TO TRAIL WITH THE 9MM AND DETERMINED THAT IT WAS A TOUCH TO BIG. THE PROCEDURE WAS COMPLETED BY TRIALING THE ACTUAL 8MM POLY. NO DELAY IN THE CASE. NO PIECES FELL INTO THE PATIENT. NO HARM TO THE PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10698242 ↗MalfunctionFeb 3, 2020Break; Break
IT WAS REPORTED THAT DURING A TKA PROCEDURE, GAP SPACE BETWEEN THE FEMORAL AND TIBIAL COMPONENT WAS TOO TIGHT. SURGEON TRIED TO FORCE THE TIBIAL INSERT TRIAL INTO SPACE, CAUSING IT TO BREAK. NO DELAY. BACKUP DEVICE AVAILABLE. NO INJURY OR IMPACT TO PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9658951 ↗MalfunctionSep 23, 2019Break; Break
IT WAS REPORTED THAT THE POLY TRIAL CRACKED DOWN THE MIDDLE. NO INJURIES OR DELAYS REPORTED. BACKUP DEVICE AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9104535 ↗MalfunctionMay 31, 2019Break; Break
IT WAS REPORTED THAT TRIALS BROKE UPON TRIALING. NO INJURY REPORTED. IS UNKNOWN IF A DELAY OCCURRED OR IF A BACKUP DEVICE WAS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8660902 ↗MalfunctionApr 15, 2019Fracture; Fracture
IT WAS REPORTED THAT THE DEVICE CRACKED DURING A CASE. THE PROCEDURE WAS COMPLETED WITH NO DELAYS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8512714 ↗MalfunctionApr 2, 2019Break; Break
IT WAS REPORTED THAT THE DEVICE BROKE DURING CASE. NO DELAY, INJURY OR AN AVAILABLE BACK UP DEVICE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8475274 ↗MalfunctionMar 26, 2019Break; Break
IT WAS REPORTED THAT DURING A TKA PROCEDURE, TIBIAL TRIAL BROKE WHILE TRIALING COMPONENTS . NO DELAY. BACKUP DEVICE AVAILABLE. NO INJURY OR IMPACT TO PATIENT REPORTED. ALL PIECES WERE RECALLED AND REMOVED FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8452411 ↗
Other TEMPLATE devices
- ATTUNE 2544-01-014Depuy (Ireland) · TEMPLATE802 reports
- ATTUNE 2545-00-546Depuy (Ireland) · TEMPLATE1 recall301 reports
- ARTICUL/EZE 2531-51-000Depuy (Ireland) · TEMPLATE323 reports
- DELTA XTEND 2307-95-000Depuy (Ireland) · TEMPLATE315 reports
- ATTUNE 2545-00-547Depuy (Ireland) · TEMPLATE1 recall231 reports
- ATTUNE 2544-00-520Depuy (Ireland) · TEMPLATE279 reports
- ATTUNE 2545-00-545Depuy (Ireland) · TEMPLATE1 recall221 reports
- ARTICUL/EZE 2530-91-000Depuy (Ireland) · TEMPLATE271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page