NA 72204925Recalls & reported problems
Smith & Nephew, Inc. · UNIVERSAL LIGHT GUIDE ORANGE 5MM X 12 FT · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00885556643747), received between May 21, 2020 and Dec 24, 2022 — classified as 1 injury, 14 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Contamination /Decontamination Problem13
- Material Separation1
- Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- No Information1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 24, 2022Output Problem; Output Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE LIGHT GUIDE HAS BEEN DISPLAYING A YELLOW-TONE IMAGE, IN ADDITION THE SCREENS INDICATED THE FOLLOWING MESSAGE "INCOMPLETE WHITE BALANCE". THERE WAS A SURGICAL DELAY OF MORE THAN 30 MINUTES IN THE PROCEDURE AND IT WAS FINISHED WITH THE SAME FAULTY DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16045841 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096459 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096435 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096424 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096366 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096355 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096327 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096326 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096304 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096232 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096199 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096194 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096158 ↗MalfunctionDec 30, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT THE "LIGHT GUIDE ORANGE (5MM X 12FT)" HAD A DARK SUBSTANCE AROUND THE CONNECTION POINT. THE CUSTOMER DISCUSSED THE INSTRUCTIONS FOR USE AND CONFIRMED THEY USING CORRECTLY 270°, 4 MINUTES, 30 COOL DOWN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11096149 ↗InjuryMay 21, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A KNEE PROCEDURE, THE LIGHT GUIDE PEELED AFTER STERILIZATION FOR THE FIRST TIME. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE. NO DELAY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10080542 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026