VISIONAIRE V0100107Recalls & reported problems

Smith & Nephew, Inc. · VISIONAIRE ADAPTIVE GUIDE DISTAL FEMUR LEGION PRIMARY · Class 2

All models of VISIONAIRE (25)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556656372), received between Jul 26, 2019 and Apr 2, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem1
  • Defective Device1
  • Failure to Align1
  • Inadequacy of Device Shape and/or Size1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Injury1
  • Laceration(s)1
  • No Consequences Or Impact To Patient1
  • No Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 2, 2024Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THAT DURING A TKA SURGERY USING VISIONAIRE FEMORAL CUTTING BLOCK, MORE BONE WAS RESECTED FROM THE FEMORAL POSTERIOR MEDIAL CONDYLE THAN PLANNED. THE PLANNED CUT WAS 9.5MM, BUT THE MEASURED RESECTION WITH A CALIPER WAS 12.5MM. THE PROCEDURE WAS RESUMED AFTER A SIGNIFICANT DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Laceration(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 19029206 ↗
  • MalfunctionJul 29, 2020Failure to Align; Failure to Align

    IT WAS REPORTED THAT DURING TKA SURGERY, THE FEMUR CUTTING GUIDE VIS ADPT GUIDE LGNP FEM WAS NOT MAKING CONTACT WITH ANTERIOR HOOK AND WAS DIFFICULT TO PIN DOWN DUE TO ROCKING. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK UP DEVICE, WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO INJURY TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10337967 ↗
  • InjuryJul 17, 2020Defective Device; Defective Device

    IT WAS REPORTED THAT DURING TKA SURGERY, THE FEMUR CUTTING GUIDE VIS ADPT GUIDE LGNP FEM DID NOT MAKE CONTACT WITH THE FEMUR ANTERIOR HOOK OR WITH DISTAL. THE SURGEON HAD TO CHANGE SURGICAL TECHNIQUE AND SWITCHED TO STANDARD INSTRUMENTS. THE PROCEDURE FINISHED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO INJURY TO THE PATIENT.

    Patient problems: Injury; No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10292542 ↗
  • MalfunctionJul 26, 2019Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT WHEN THE SURGEON ATTEMPTED TO PLACE THE FEMORAL CUTTING BLOCK ON THE PATIENT, HE WAS HAVING BLOCK FIT ISSUES. THE BLOCK DID NOT APPEAR TO MATCH THE ANTERIOR CORTEX OF THE PATIENT'S FEMORAL BONE ANATOMY. THE SURGEON DID NOT USE THE CUTTING GUIDE AND WENT WITH STANDARD INSTRUMENTS. NO DELAYS REPORTED. NO PATIENT INJURIES REPORTED. IT IS UNKNOWN IF AN S&N BACKUP DEVICE WAS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8834206 ↗

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page