Skip to content

OR3O Dual Mobility 71353302Recalls & reported problems

Smith & Nephew, Inc. · OR3O DUAL MOBILITY TRIAL LINER 38/50 · Class 2

All models of OR3O Dual Mobility (30)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556710456), received between Nov 10, 2020 and Dec 29, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING AN ACETABULAR REVISION SURGERY, AT THE TIME OF PERFORMING TESTS WITH THE OR3O DUAL MOBILITY TRIAL LINER 38/50, IT WAS ATTACHED TO THE ACETABULAR CUP USING THE INTERNAL RETAINING SCREW. THE RELEVANT TESTS WERE COMPLETED AND UPON DISASSEMBLING IT FROM THE CUP, THE RETAINING SCREW CAME LOOSE, COMPLETELY DETACHING AND PREVENTING IT FROM BEING REATTACHED. THE SCREW DID NOT FELL INSIDE THE PATIENT. THIS OCCURRED AFTER ALL NECESSARY TESTS FOR THE IMPLANTATION OF THE FINAL DEVICE HAD ALREADY BEEN PERFORMED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23917851 ↗
  • MalfunctionAug 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING THR SURGERY, IT WAS NOTICED THAT, ONE (1) OR3O DUAL MOBILITY TRIAL LINER 38/50 BROKE DURING USE INSIDE THE PATIENT. NO ONE WAS HARMED. THE SCREW INSIDE THE LINER TRIAL SEEMED TO FALL APART WHEN THEY WERE TRYING TO SCREW IT INTO THE SHELL. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, CHANGING SURGICAL TECHNIQUE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19991496 ↗
  • MalfunctionOct 5, 2022Break; Break

    IT WAS REPORTED THAT IN ONE (1) OR3O DUAL MOBILITY TRIAL LINER 38/50, A METAL ROUND PIECE BROKE AND FELL OUT, NOW SCREW IS NOT SECURE. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15544181 ↗
  • MalfunctionJul 28, 2022Break; Break

    IT WAS REPORTED THAT, AN OR3O DUAL MOBILITY TRIAL LINER 38/50 WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15125914 ↗
  • MalfunctionDec 20, 2021Break; Break

    IT WAS REPORTED THAT DURING A THR, THE OR3O DUAL MOBILITY TRIAL LINER 38/50 BROKE OUTSIDE THE PATIENT. NO PIECES FELL INSIDE. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13048689 ↗
  • InjuryNov 10, 2020Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE SCREW AND RING OF THE OR3O DUAL MOBILITY TRIAL LINER 38/50 BECAME DISASSOCIATED FROM PLASTIC DURING TRIALING. INSTRUMENTS INSIDE PATIENT. ALL PIECES RECOVERED. PROCEDURE FINISHED WITH SAME DEVICE. DELAY OF FIVE MINUTES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10816925 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page