Skip to content

AETOS Shoulder System SCR-0990-862Recalls & reported problems

Smith & Nephew, Inc. · T-20 DRIVER FOR 4.5MM SCREWS · Class 2

All models of AETOS Shoulder System (44)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885556812235), received between Jul 1, 2024 and Dec 22, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A AETOS SHOULDER SURGERY, THE TIP OF ONE (1) T-20 DRIVER 4.5MM PERIPHERAL SCREWS BROKE OFF AFTER PUTTING LAST SCREW IN. DRIVER TIP DID NOT FALL INTO PATIENT. PIECE WAS RETRIEVED BY HAND. THE PROCEDURE WAS ALREADY COMPLETED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872356 ↗
  • MalfunctionNov 14, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TSA SURGERY, ONE (1) T-20 DRIVER 4.5MM PERIPHERAL SCREWS IS BROKEN OFF. NO PIECE FELL INSIDE PATIENT'S WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23556927 ↗
  • MalfunctionAug 2, 2025Fracture; Material Deformation; Fracture; Material Deformation

    IT WAS REPORTED THAT DURING A REVERSE SHOULDER ARTHROPLASTY, THE TIP OF THE T-20 DRIVER 4.5MM PERIPHERAL SCREWS SNAPPED OFF WHILE TIGHTENING THE PERIPHERAL LOCKING SCREW, THE BROKEN TIP WAS REMOVED BY SUCTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22694155 ↗
  • MalfunctionSep 10, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TSA SURGERY, THE DISTAL TIP OF ONE (1) T-20 DRIVER 4.5MM PERIPHERAL SCREWS WAS TWISTED OFF WHILE USING ONE (1) GLENOSPHERE REMOVAL TOOL 42. THE FRACTURED PIECE WAS CONTAINED WITHIN THE REMOVAL TOOL, AND IT WAS RETRIEVED FROM THE TOOL ON THE BACK TABLE AFTER SURGERY. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20186654 ↗
  • MalfunctionJul 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A REVERSE SHOULDER ARTHROPLASTY, UPON TIGHTENING A LOCKING SCREW INTO A BASEPLATE, THE TIP OF THE T-20 DRIVER 4.5MM PERIPHERAL SCREWS SNAPPED. THE PIECE THAT FELL INSIDE THE PATIENT WAS REMOVED USING SUCTION. THE PROCEDURE WAS PERFORMED, AFTER NO DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19644938 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page