Skip to content

AETOS Shoulder System STEM-0990-100-002Recalls & reported problems

Smith & Nephew, Inc. · META HUMERAL STEM SIZE 2 SMALL · Class 2 · Implant

All models of AETOS Shoulder System (44)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556814222), received between Mar 25, 2024 and Sep 5, 2024 — classified as 2 injuries.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Bone Fracture(s)1
  • Medical device site infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, DURING AN AETOS SURGERY FOR A RIGHT ROTATOR CUFF REPAIR, THE GREATER TUBEROSITY FRACTURE DURING SURGERY. THIS WAS TREATED WITH A CERCLAGE FIXATION. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Bone Fracture(s)

    MAUDE report 20152384 ↗
  • InjuryMar 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER TSA SURGERY HAD BEEN PERFORMED ON (B)(6) 2024, THE PATIENT EXPERIENCED INFECTION. THIS ADVERSE EVENT WAS SOLVED BY REVISION SURGERY ON (B)(6) 2024, IN WHICH THE META HUMERAL STEM SIZE 2 S WAS EXPLANTED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Medical device site infection

    MAUDE report 18969265 ↗

Other shoulder prosthesis, reverse configuration devices

All shoulder prosthesis, reverse configuration devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page