Skip to content

AETOS Shoulder System GBP-0990-502-15FRecalls & reported problems

Smith & Nephew, Inc. · 15 DEG AUGMENT BASEPLATE FULL WEDGE · Class 2 · Implant

All models of AETOS Shoulder System (44)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556814390), received between Jan 31, 2024 and Mar 6, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Fracture1
  • Material Deformation1
  • Unintended Movement1

Patient problems

  • Medical device site infection2
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER AN AETOS TSA SURGERY HAD BEEN PERFORMED ON (B)(6) 2023, THE PATIENT EXPERIENCED SIGNS OF INFECTION. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2026, IN WHICH THE 15 DEG AUGMENT BASEPLATE FULL WEDGE WAS EXPLANTED, AND STRYKER IMPLANTS WERE IMPLANTED IN A SINGLE-STAGE REVISION. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Medical device site infection

    MAUDE report 24531115 ↗
  • InjuryApr 26, 2025Adverse Event Without Identified Device or Use Problem; Unintended Movement; Adverse Event Without Identified Device or Use Problem; Unintended Movement

    IT WAS REPORTED THAT, AFTER AN RSA PERFORMED ON (B)(6) 2024, THE PATIENT PRESENTED SHOULDER PAIN. THE SURGEON SUSPECTED THAT ONE OF THE PERIPHERAL SCREWS HAS BROKEN AND THE BASEPLATE MAY BE LOOSE. HE ALSO SUSPECTED THAT THE POLY LINER HAS DISASSOCIATED FROM THE STEM, AS WELL AS A POSSIBLE INFECTION. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2025 TO EXPLANT THE PREVIOUS DEVICES; THE PATIENT WAS TREATED WITH ANTI-BACTERIAL RINSE FOLLOWING THE EXPLANATION. NO OTHER SIGNS OF INFECTION WERE VISUALLY PRESENT, AND THE SURGEON PROCEEDED TO IMPLANT A TORNIER PYROCARBON HEMI CONSTRUCT. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Medical device site infection; Pain

    MAUDE report 21915681 ↗
  • MalfunctionJan 31, 2024Fracture; Material Deformation; Fracture; Material Deformation

    IT WAS REPORTED THAT DURING AN AETOS REVERSE TOTAL SHOULDER, THE HEAD OF THE 4.5MM CENTER SCREW 40MM BROKE WHILE TRYING TO ADVANCE THE SCREW, LEAVING IT TOO PROUD TO ENGAGE THE GLENOSPHERE SET SCREW. (B)(6) DR. MADE THE DECISION TO REMOVE THE 15 DEG AUGMENT BASEPLATE FULL WEDGE AND IN THE PROCESS OF REMOVING IT, THE THREADS OF THE BASEPLATE INSERTER BROKE OFF IN THE CENTER POST OF THE BASEPLATE. THE THREAD INSIDE THE BASEPLATE THAT GUIDES AND CAPTURES THE CENTER SCREW ALSO BROKE FROM INSIDE THE POST, LEAVING THE CENTER SCREW (WITHOUT THE SCREW HEAD) WITH THE CAPTURE THREAD STILL INSIDE THE VAULT OF THE GLENOID. (B)(6) DR. ENGAGED THE INSERTER ONTO THE BASEPLATE AND TRIED BACKSLAPPING THE INSERTER HANDLE AND WIGGLING THE HANDLE/BASEPLATE TO DISENGAGE THE BASEPLATE FROM THE NATIVE GLENOID. AFTER ALSO USING AN OSTEOTOME TO TRY TO LEVER THE BASEPLATE OFF THE GLENOID, HE WAS STUCK WITH THE BASEPLATE MAY A FEW MILLIMETERS OFF THE GLENOID. THINKING THE CENTER SCREW WAS HOLDING HIM UP, HE TRIED TO SPIN THE BASEPLATE COUNTERCLOCKWISE TO TRY TO UNSCREW THE BASEPLATE OFF THE CENTER SCREW. AFTER GOING BACK AND FORTH FOR A LONG TIME, THE SCREW THREADS OF THE MACHINED, BASEPLATE & GLENOSPHERE IN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18614323 ↗

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page