Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556862346), received between Jul 11, 2023 and Jul 11, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 11, 2023Break; Break

    IT WAS REPORTED THAT DURING TOTAL SHOULDER ARTHROPLASTY SURGERY, THE ATTACHMENT POINT OF ONE (X1) FIN-LOCK SIZER. M AND ONE (X1) FIN-LOCK STEPPED SIZER. M+3 BROKE. THE SCRUB TECH HAD TIGHTENED THE FITTING BETWEEN THE TWO BECAUSE THE SIZE KEPT FALLING OFF. ONCE THE SURGEON TRIED TO PLACE THE FIN-LOCK SIZER. M INSIDE THE SHOULDER, IT BROKE AT THAT ATTACHMENT POINT. IT BROKE IN TO TWO PIECES, THE MAIN BODY OF THE SIZER WHICH WAS EXTRACTED BY THE SURGEON AND THE ATTACHMENT POINT WAS RETAINED IN THE INSERTION HANDLE. WHEN THE FIRST SIZER BROKE, THEY TRIED WITH A FIN-LOCK STEPPED SIZER. M+3 AND EXPERIENCED THE SAME ISSUE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17302427 ↗