ENGAGE 1-10011-200Recalls & reported problems
Smith & Nephew, Inc. · TIBIAL ANCHOR STEM SIZE 3-4 · Class 2 · Implant
All models of ENGAGE (80)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556873090), received between Nov 14, 2022 and Oct 16, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 2, 2023Class IIOpen, Classified
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Z-1671-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture2
- Difficult to Insert1
- Material Deformation1
Patient problems
- Pain2
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 16, 2023Fracture; Fracture
IT WAS REPORTED THAT, AFTER A KNEE SURGERY HAD BEEN PERFORMED A YEAR AGO, THE PATIENT EXPERIENCED PAIN. IMAGES WERE TAKEN AND IT WAS FOUND THAT THE ENGAGE TIBIAL ANCHOR STEM SZ 3-4 FRACTURED AND THE ENGAGE POROUS TIBIAL TRAY SZ 4-LT MED SUBSIDED. THIS ADVERSE EVENT WAS SOLVED BY REVISION SURGERY ON (B)(6) SEP, IN WHICH THE COMPONENTS WERE REMOVED AND A PRIMARY KNEE WAS IMPLANTED WITH TJO. IT IS UNKNOWN THE CURRENT HEALTH STATUS OF PATIENT.
Patient problems: Failure of Implant; Pain
MAUDE report 17939800 ↗InjuryMay 26, 2023Fracture; Fracture
IT WAS REPORTED THAT, EIGHT MONTHS AFTER UNICOMPARTMENTAL KNEE ARTHROPLASTY WAS PERFORMED ON (B)(6) 2020, THE PATIENT EXPERIENCED PAIN. A POST-OPERATIVE X-RAY SHOWED RADIOLUCENCY AROUND THE ANCHOR AND FRACTURE OF THE ANCHOR. BASED ON THE X-RAYS PROVIDED, THE IMPLANT WAS MALPOSITIONED WITH THE TRAY NOT FULLY SEATED ANTERIORLY. LOADING OF THE IMPLANT DURING NORMAL ACTIVITY WOULD RESULT IN HIGHER THAN NORMAL FORCES GENERATED ON THE ANCHOR COMPONENT, CAUSING A LEVERING EFFECT. THIS ADVERSE EVENT WAS TREATED WITH REVISION SURGERY TO TKA ON (B)(6) 2021.
Patient problems: Pain
MAUDE report 17010869 ↗MalfunctionNov 14, 2022Difficult to Insert; Material Deformation; Difficult to Insert; Material Deformation
IT WAS REPORTED THAT, DURING A TKA PROCEDURE, SURGEON TRIED TO IMPLANT THE TIBIAL ANCHOR STEM SIZE 3-4 AND IT WOULD NOT FIT AND AS THE DOCTOR PUSHED THE INSERTER HARDER, THE ANCHOR STEM CAUGHT THE EDGE OF SOMETHING AND BEGAN TO BEND. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT HARMED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15791785 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-1673-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)
- Z-1670-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
- Z-1672-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1
Other Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer devices
- ENGAGE 1-10003-410Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall10 reports
- ENGAGE 1-10012-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall6 reports
- ENGAGE 1-10003-600Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall4 reports
- ENGAGE 1-10003-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10012-350Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10003-310Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall2 reports
- ENGAGE 1-10012-200Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall2 reports
- ENGAGE 1-10003-200Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page