Skip to content

N/A (REF 401860)Recalls & reported problems

Stryker Sustainability Solutions, Inc. · Diagnostic EP Catheter CRD-2 Curve, 4 electrodes, 5mm spacing, 5F · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885825007034), received between Apr 13, 2017 and Nov 1, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Knotted” (1 report).

29 other recalls cover devices cleared under the same FDA submission (K012708); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Knotted1
  • Use of Device Problem1

Patient problems

  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 1, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A CATHETER GUIDE WIRE BROKE AND WAS LEFT IN THE PATIENT. AS REPORTED, PLAIN FILM AND CT IMAGING REVEALED THE LOCATION OF THE WIRE. AN ADDITIONAL PROCEDURE WAS REQUIRED TO SUCCESSFULLY REMOVE THE WIRE. THERE WAS NO OTHER MEDICAL INTERVENTION REPORTED AND THE COMPLAINANT WAS NOT AWARE OF ANY EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20592856 ↗
  • MalfunctionApr 13, 2017Knotted

    IT WAS REPORTED THE CATHETER KNOTTED DURING SURGERY. THE DEVICE AND SHEATH WERE PULLED OUT OF THE PATIENT WITH THE CATHETER STILL KNOTTED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME WAS A FEW MINUTES TO REMOVE THE CATHETER FROM THE PATIENT. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6491186 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K012708. One submission often covers many models, so check the product and lot details in each record.

  • Z-2178-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2185-2026 · Apr 10, 2026 · Open, Classified
    CS Uni-Directional w Auto ID, Product Number D135304; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2186-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401575; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2182-2026 · Apr 10, 2026 · Open, Classified
    BARD EP XT Steerable, Product Number 200794; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2189-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2174-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2188-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2177-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic Tip Steerable, Product Number 6DYNTP001;
  • Z-2192-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401905; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2199-2026 · Apr 10, 2026 · Open, Classified
    Medtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2176-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2179-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2190-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2198-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable BDB, Product Number 401918, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2184-2026 · Apr 10, 2026 · Open, Classified
    CS Uni-Directional w Auto ID, Product Number D135303; REPROCESSED ELECTROPHYSIOLOGY CATHETER

…and 14 more.

Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices

All CATHETER, RECORDING, ELECTRODE, REPROCESSED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.