N/A (REF 401904)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Steerable Diagnostic EP Catheter Super Large Curl, 7F · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885825007263), received between Sep 16, 2019 and Dec 22, 2025 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K052603); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 10, 2026Class IIOpen, Classified
Incomplete seals on sterile product
Z-2191-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Use of Device Problem3
- Mechanical Problem2
- Positioning Problem2
- Deformation Due to Compressive Stress1
- Electrical /Electronic Property Problem1
- Mechanical Jam1
- Poor Quality Image1
- Structural Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Consequences Or Impact To Patient1
- Tricuspid Valve Insufficiency/ Regurgitation1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025Electrical /Electronic Property Problem; Mechanical Problem; Use of Device Problem; Electrical /Electronic Property Problem; Mechanical Problem; Use of Device Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE LIVEWIRE DUODECA CATHETER. THE PINS DID NOT HAVE SIGNAL AND THE CATHETER WAS DIFFICULT TO STEER AND POSITION PROPERLY. AS REPORTED, THIS OCCURRED DURING AN AFIB ABLATION AND ADDED TO THE FLUORO TIME AND THE TIME THE PATIENT WAS UNDER GENERAL ANESTHESIA, AFFECTING BOTH THE PATIENT AND THE STAFF. THE CATHETER WAS REPLACED WITH AN OEM DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS THE TIME OF GRABBING THE OEM DEVICE AND USING THAT DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23870700 ↗MalfunctionDec 22, 2025Poor Quality Image; Use of Device Problem; Poor Quality Image; Use of Device Problem
IT WAS REPORTED THAT THERE WAS AN ISSUE WITH THE LIVEWIRE DUODECA CATHETER. THE PINS DID NOT HAVE SIGNAL AND THE CATHETER WAS DIFFICULT TO STEER AND POSITION PROPERLY. AS REPORTED, THIS OCCURRED DURING AN AFIB ABLATION AND ADDED TO THE FLUORO TIME AND THE TIME THE PATIENT WAS UNDER GENERAL ANESTHESIA, AFFECTING BOTH THE PATIENT AND THE STAFF. THE CATHETER WAS REPLACED WITH AN OEM DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS THE TIME OF GRABBING THE OEM DEVICE AND USING THAT DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23870697 ↗MalfunctionApr 21, 2025Positioning Problem
DURING ATRIAL FLUTTER ABLATION PROCEDURE, IT WAS REPORTED THAT LIVEWIRE COULD NOT EASILY BE PLACED IN DESIRED LOCATION. AT ONE POINT, DIFFERENT SHEATHS WERE USED AS DILATORS, EVENTUALLY IT WAS DETERMINED THAT THE CATHETER PLACEMENT WAS NOT ADEQUATE, HAD TO BE REMOVED, AND DR [REDACTED] ASKED FOR ANOTHER LIVEWIRE. LIVEWIRE IN REPORT WAS REMOVED AND ANOTHER PLACED WITHOUT ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873148 ↗MalfunctionMar 31, 2025Deformation Due to Compressive Stress; Mechanical Jam; Deformation Due to Compressive Stress; Mechanical Jam
IT WAS REPORTED THE CATHETER HAD A CURVE ON THE WIRE WHICH CREATED A KINK. AS REPORTED, STAFF WERE NOT ABLE TO CANNULATE THE CORONARY SINUS. THE DEVICE CAME OUT THE BOX WITH THE ISSUE AND A NEW NON-REPROCESSED REPLACEMENT WAS USED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS TEN MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21733476 ↗MalfunctionNov 19, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED CATHETER WOULD NOT DEFLECT. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12846794 ↗MalfunctionDec 16, 2020Structural Problem; Structural Problem
IT WAS REPORTED THAT THE CATHETER WAS 'TOO FLIMSY TO MANIPULATE'. IT WAS REMOVED AND REPLACED. THE CATHETER WAS FLIMSY AND LIKE A NOODLE WHEN MANIPULATED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME LENGTH IS UNKNOWN. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11016476 ↗InjurySep 25, 2020Use of Device Problem; Use of Device Problem
IT WAS REPORTED DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE, THE CATHETER GOT STUCK IN THE RIGHT VENTRICULAR APEX WHILE ADVANCING TO THE CORONARY SINUS CAUSING SOME TRICUSPID REGURGITATION. A SHEATH HAD TO BE INSERTED TO REMOVE THE CATHETER. THE MEDWATCH FORM SELECTED "OTHER SERIOUS (IMPORTANT MEDICAL EVENTS)", HOWEVER NO FURTHER DETAILS WERE PROVIDED FROM THE CUSTOMER ABOUT THE EVENT. MULTIPLE ATTEMPTS WERE MADE FOR ADDITIONAL INFORMATION, HOWEVER NONE WERE PROVIDED. THERE WERE NO ADVERSE CONSEQUENCES OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Tricuspid Valve Insufficiency/ Regurgitation
MAUDE report 10590447 ↗MalfunctionSep 16, 2019Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE REPROCESSED LIVE WIRE ELECTROPHYSIOLOGY CATHETER WOULD NOT STEER. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL AT 5 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9070488 ↗
FDA clearance and classification
Cleared through 510(k) K052603 on Nov 16, 2005 (decision: Substantially Equivalent). Submitted by Alliance Medical Corp. as “Reprocessed Daig Livewire Electrophysiology Catheters”. The FDA review took 56 days. FDA record ↗
The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K052603. One submission often covers many models, so check the product and lot details in each record.
- Z-2185-2026 · Apr 10, 2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135304; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2188-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2190-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2196-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401923; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2189-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2180-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2184-2026 · Apr 10, 2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135303; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2176-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2187-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401581; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2183-2026 · Apr 10, 2026 · Open, ClassifiedBW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2175-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic Tip Steerable, Product Number 200131; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2198-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable BDB, Product Number 401918, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2192-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401905; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2193-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401908; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2178-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
…and 10 more.
Common questions about the N/A
Has the N/A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2191-2026 (Class II), began on Apr 10, 2026: Incomplete seals on sterile product. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the N/A?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Sep 16, 2019 and Dec 22, 2025. In those reports, the most common specific device problems are “Use of Device Problem” (3), “Mechanical Problem” (2) and “Positioning Problem” (2); the most common patient problem is “Tricuspid Valve Insufficiency/ Regurgitation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A?
None of the reports are classified as deaths, 1 as an injury and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A FDA approved?
It was cleared through 510(k) K052603 on Nov 16, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the N/A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the N/A, and is it still sold?
The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices
- MEDLINE 401904RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls1 report
- MEDLINE 401582RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls0 reports
- MEDLINE 401474RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls1 report
- MEDLINE 81174RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401443RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE BD710FJ282RRHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401652RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401451RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page