N/A (REF 401932)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Steerable Diagnostic EP Catheter Super Large Curl, 7F · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 00885825007355), received between Dec 21, 2016 and Dec 8, 2025 — classified as 1 injury, 20 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
25 other recalls cover devices cleared under the same FDA submission (K052603); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 10, 2026Class IIOpen, Classified
Incomplete seals on sterile product
Z-2197-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation8
- Use of Device Problem5
- Mechanical Problem4
- Difficult to Remove2
- Break1
- Device Operates Differently Than Expected1
- Difficult To Position1
- Entrapment of Device1
Patient problems
- No Consequences Or Impact To Patient12
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE DEVICE DID NOT STEER TIGHT ENOUGH BY THE PHYSICIAN. AS REPORTED, THE DEVICE HAD MULTIPLE ATTEMPTS TO NAVIGATE FEMORAL VEIN TO CS THAT WERE UNSUCCESSFUL. ICE WAS INTRODUCED TO HELP GUIDE THE DEVICE TO THE PROPER LOCATION. ATTEMPTS WERE UNSUCCESSFUL. UPON REPLACEMENT WITH ANOTHER STRYKER DUADECA, ACCESS TO THE CS WAS GAINED UPON THE FIRST ATTEMPT. THE DEVICE WAS REPLACED WITH A STRYKER DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 3-5 MINUTES AS ACCESS WAS ATTEMPTED AND REPLACEMENT WAS INTRODUCED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23751090 ↗MalfunctionMar 10, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT JUST CUT TIP - BROKEN. THE COMPLAINANT IS NOT AWARE OF ANY PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 21562561 ↗MalfunctionMar 10, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THERE WAS POOR STEERABILITY IN THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21562229 ↗MalfunctionMar 10, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THERE WAS POOR STEERABILITY IN THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21562050 ↗MalfunctionAug 14, 2024Mechanical Problem; Use of Device Problem; Mechanical Problem; Use of Device Problem
IT WAS REPORTED FOR THE CATHETER THAT THE ITEM DIDN'T FUNCTION PROPERLY MECHANICALLY DURING PROCEDURE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 2 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19986568 ↗MalfunctionDec 10, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THE PRODUCT WAS BENT AND UNABLE TO HOLD SHAPE. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12980879 ↗MalfunctionSep 14, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THE PHYSICIAN WAS ATTEMPTING TO USE A REPROCESSED ABBOTT LIVEWIRE DIAGNOSTIC EP CATHETER DURING THE PROCEDURE, BUT REALIZED THAT THE CATHETER WAS NOT DEFLECTING CORRECTLY, AND THERE WAS A KINK/BEND BETWEEN SOME OF THE ELECTRODES ON THE CATHETER THAT PREVENTS THE CATHETER FROM RETURNING TO A STRAIGHT POSITION. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS LESS THAN 5 MINUTES WHILE STAFF WENT TO RETRIEVE A NEW REPLACEMENT CATHETER FROM THE STOREROOM. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12467096 ↗MalfunctionJul 29, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED DURING A SVT ABLATION PROCEDURE PHYSICIAN ATTEMPTED TO USE A REPROCESSED ABBOTT LIVEWIRE DUODECA WITH SPACING AND NOTED THAT THE CATHETER WAS NOT CURVING CORRECTLY AND HAD A BEND/KINK BETWEEN A SET OF ELECTRODES THAT CAUSED A DISSATISFYING CURVE SHAPE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 3 TO 5 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12250406 ↗MalfunctionMar 3, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER BECAME KINKED AS IT WAS BEING ADVANCED UP TO THE HEART. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9780787 ↗MalfunctionMar 3, 2020Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER BENT WHEN THE PHYSICIAN ATTEMPTED TO DEFLECT IT. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER REVEALED THE ELECTROPHYSIOLOGY CATHETER HAD A KINK AMIDST THE ELECTRODES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS LESS THAN 30 MINUTES TO REPLACE THE DEVICE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9780786 ↗InjuryJan 15, 2020Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER STEERING MECHANISM FAILED AND THE DOCTOR HAD DIFFICULTY REMOVING THE DEVICE FROM THE PATIENT'S BODY. THE CATHETER WAS STUCK IN THE PATIENT'S HEART AND WAS NOT DEFLECTING. TWO ADDITIONAL PHYSICIANS AND TWO ADDITIONAL DEVICES WERE REQUIRED TO REMOVE THE ELECTROPHYSIOLOGY CATHETER. THERE WAS A DELAY IN THE CASE OF APPROXIMATELY 45 MINUTES TO REMOVE AND REPLACE THE DEVICE. THE PATIENT WAS FINE AND THERE WAS NO PATIENT INJURY. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9591071 ↗MalfunctionJul 12, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER WAS FRACTURED; THERE WAS A MANEUVERABILITY ISSUE PRIOR TO THE PROCEDURE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8786003 ↗MalfunctionJul 2, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER WAS KINKED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS 5 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8755603 ↗MalfunctionMar 5, 2019Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING THE CASE, THE ELECTROPHYSIOLOGY CATHETER BECAME EXTREMELY BENT AND WOULD NOT READ PROPERLY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8392790 ↗MalfunctionFeb 21, 2019Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER KINKED IN WITHIN THE PATIENT'S HEART. THEY HAD TO MAKE SEVERAL MANEUVERS TO UNKINK AND REMOVE IT FROM THE PATIENT'S HEART. THE DR. USED ANOTHER DEVICE TO FINISH THE PROCEDURE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8358108 ↗MalfunctionFeb 21, 2019Difficult to Remove; Material Deformation; Difficult to Remove; Material Deformation
IT WAS REPORTED THAT AN ELECTROPHYSIOLOGY CATHETER WAS KINKED, AND THE KINK HAD TO BE MANIPULATED TO REMOVE THE DEVICE FROM THE PATIENT'S BODY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS APPROXIMATELY ONE MINUTE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8357834 ↗MalfunctionFeb 14, 2019Positioning Problem; Positioning Problem
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER HAD A FAULTY DEFLECTION MECHANISM. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8339203 ↗MalfunctionDec 12, 2018Material Deformation; Material Deformation
IT WAS REPORTED THE TIP OF THE ELECTROPHYSIOLOGY CATHETER BECAME BENT DURING THE PROCEDURE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 5 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8155120 ↗MalfunctionSep 27, 2017Difficult To Position; Difficult To Position
IT WAS REPORTED THERE WAS A STEERING ISSUE WITH THE ELECTROPHYSIOLOGY CATHETER. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6898975 ↗MalfunctionFeb 9, 2017Break
IT WAS REPORTED THE CATHETER WAS MISSHAPED/BROKEN. THERE WAS A MINIMAL DELAY TO REPLACE THE DEVICE. IT WAS STATED BY THE CUSTOMER THE DOCTOR MARKED "POSSIBLE PATIENT HARM" ON HIS SHEET. NO ADDITIONAL DETAILS WERE PROVIDED AS THAT WAS ALL THE INFORMATION THE CUSTOMER HAD. THERE WAS NO MEDICAL INTERVENTION REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6315884 ↗MalfunctionDec 21, 2016Device Operates Differently Than Expected
IT WAS REPORTED THAT THE DOCTOR HAD TROUBLE STEERING THE DEVICE INTO THE CORONARY SINUS DURING AN AFIB PROCEDURE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6194172 ↗
FDA clearance and classification
Cleared through 510(k) K052603 on Nov 16, 2005 (decision: Substantially Equivalent). Submitted by Alliance Medical Corp. as “Reprocessed Daig Livewire Electrophysiology Catheters”. The FDA review took 56 days. FDA record ↗
The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K052603. One submission often covers many models, so check the product and lot details in each record.
- Z-2179-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2194-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401914; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2174-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2181-2026 · Apr 10, 2026 · Open, ClassifiedBARD EP XT Decapolar Steerable, Product Number 201007; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2183-2026 · Apr 10, 2026 · Open, ClassifiedBW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2185-2026 · Apr 10, 2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135304; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2196-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401923; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2188-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2184-2026 · Apr 10, 2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135303; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2190-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2187-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401581; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2189-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2178-2026 · Apr 10, 2026 · Open, ClassifiedBARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2199-2026 · Apr 10, 2026 · Open, ClassifiedMedtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2195-2026 · Apr 10, 2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401915; REPROCESSED ELECTROPHYSIOLOGY CATHETER
…and 10 more.
Common questions about the N/A
Has the N/A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2197-2026 (Class II), began on Apr 10, 2026: Incomplete seals on sterile product. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the N/A?
The FDA's MAUDE database lists 21 adverse event reports for this device, received between Dec 21, 2016 and Dec 8, 2025. In those reports, the most common specific device problems are “Material Deformation” (8), “Use of Device Problem” (5) and “Mechanical Problem” (4). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A?
None of the reports are classified as deaths, 1 as an injury and 20 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A FDA approved?
It was cleared through 510(k) K052603 on Nov 16, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the N/A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the N/A, and is it still sold?
The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices
- MEDLINE 401904RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls1 report
- MEDLINE 401582RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls0 reports
- MEDLINE 401474RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls1 report
- MEDLINE 81174RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401443RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE BD710FJ282RRHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401652RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401451RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page