N/A (REF 7003D)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Steerable Diagnostic EP Catheter , STANDARD, 10 electrodes, 2.5-5-2.5mm spacing, 6F · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00885825008710), received between Jun 18, 2018 and Mar 11, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation2
- Device Damaged Prior to Use1
- Difficult To Position1
- Inadequacy of Device Shape and/or Size1
- Kinked1
- Material Integrity Problem1
- Mechanical Problem1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. THERE WAS NO PATIENT INJURY AND THE COMPLAINANT IS NOT AWARE OF ANY MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21572190 ↗MalfunctionJun 3, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED THE POLARIS EP CATHETER WAS NOTED TO BE DAMAGED WHEN OPENED. THE CATHETER WAS BENT THEREFORE NOT ABLE TO BE USED AND ANOTHER ONE HAD TO BE OPENED. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11935589 ↗MalfunctionMay 20, 2019Material Deformation; Positioning Problem; Material Deformation; Positioning Problem
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER CAME OUT OF THE PACKAGE WITH A CURLED END. THE STAFF TRIED USING IT IN THE PATIENT, BUT THEY COULDN'T GET IT TO WORK - THERE WAS A DEFLECTION ISSUE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8625897 ↗MalfunctionFeb 11, 2019Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED THAT THERE WAS A DAMAGED ELECTROPHYSIOLOGY CATHETER. THE FLEXOR WAS DAMAGED AT THE TIME THE PACKAGE WAS OPENED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THE COMPLAINANT WAS NOT AWARE OF THE DURATION OF EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8326167 ↗MalfunctionOct 4, 2018Material Deformation; Material Deformation
IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER CAME IN BENT AND DID NOT WORK WHEN IT WAS PLUGGED IN. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7935497 ↗MalfunctionSep 13, 2018Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THE CATHETER DOES NOT CURVE CORRECTLY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7874622 ↗MalfunctionJun 18, 2018Kinked; Difficult To Position; Kinked; Difficult To Position
IT WAS REPORTED THAT THE TIP OF THE ELECTROPHYSIOLOGY CATHETER WAS KINKED AND COULD NOT GET IN PLACE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7612239 ↗
FDA clearance and classification
Cleared through 510(k) K052414 on Nov 16, 2005 (decision: Substantially Equivalent). Submitted by Alliance Medical Corp. as “Reprocessed Electrophysiology Catheter”. The FDA review took 75 days. FDA record ↗
The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A
Has the N/A been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885825008710), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the N/A?
The FDA's MAUDE database lists 7 adverse event reports for this device, received between Jun 18, 2018 and Mar 11, 2025. In those reports, the most common specific device problems are “Material Deformation” (2), “Device Damaged Prior to Use” (1) and “Difficult To Position” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A?
None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the N/A FDA approved?
It was cleared through 510(k) K052414 on Nov 16, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the N/A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the N/A, and is it still sold?
The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices
- MEDLINE 401904RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls1 report
- MEDLINE 401582RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls0 reports
- MEDLINE 401474RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls1 report
- MEDLINE 81174RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401443RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE BD710FJ282RRHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401652RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401451RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.