IPure B3PC (clear), +19.50DRecalls & reported problems

Hoya Medical Singapore Pte. Ltd. · Foldable, posterior chamber intraocular lens (IOL), preloaded in an injector that automatically folds the lens as it is advanced, prior to insertion. Single use. · Class 3 · Implant

All models of IPure (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886158015772), received between Nov 13, 2020 and Jul 8, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Crack1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2025Crack; Crack

    EVENT OCCURRED IN UNITED STATES. DAMAGED OPTIC AFTER IMPLANTATION. CRACKED OR BROKEN OPTIC (MIDDLE). THE PATIENT IS CURRENTLY STABLE. THERE WAS NO HARM TO THE PATIENT. AN IRREGULAR MAIN WOUND OCCURRED DURING THE CASE, BUT IT WAS SEALED WELL WITH SUTURE. THE PHYSICIAN REMOVED THE BISECTED OPTIC FROM THE ANTERIOR CHAMBER. THERE WILL BE NO ONGOING TREATMENT. SUTURES WILL BE REMOVED WHEN WOUND IS STABLE. IOL WAS EXPLANTED ON (B)(6) 2025. PROBLEM CODE: A0404 - CRACK.

    Patient problems: Failure of Implant

    MAUDE report 22446902 ↗
  • InjuryNov 13, 2020Break; Break

    IOL GOT STUCK IN THE NOZZLE DURING INSERTION. USER ERROR AS TECH PRESSED SLIDER FORWARD SHE LIKELY DISPUTED BODY OF INJECTOR CAUSING LENS TO SPLIT IN HALF. SURGEON INSERTED SPLIT LENS, WIDENED INCISION AND REMOVED IOL WITH NO INJURY TO PATIENT. PATIENT IMPACT: INTRA-OPERATIVE EXPLANT INCREASED INCISION SIZE; PRODUCT REPLACED WITH ANOTHER LENS IMMEDIATELY DURING SURGERY; INCREASED INCISION REQUIRED SUTURES TO CLOSE; PATIENT HEALTH NOT IMPACTED; PATIENT HAS RECOVERED AND IS FINE; NO PERMANENT OR NEGATIVE IMPACT ON PATIENT HEALTH EXPECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10837781 ↗