Hoya Medical Singapore Pte. Ltd. · Foldable, posterior chamber intraocular lens (IOL), preloaded in an injector that automatically folds the lens as it is advanced, prior to insertion. Single use. · Class 3 · Implant

All models of IPure (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00886158017233), received between Mar 23, 2022 and Jun 12, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Break1
  • Sharp Edges1

Patient problems

  • Capsular Bag Tear1
  • Failure of Implant1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    EVENT OCCURRED IN USA. DAMAGED HAPTIC AFTER IMPLANTATION. LEADING; TIP; DAMAGED HAPTIC; HAPTIC DOES NOT FOLD AS PER IFU; HAPTIC STUCK TO OPTIC TEMPORARILY~ FINALLY HAPTIC IS SEPARATED INCREASED INCISION SIZE; PRODUCT REPLACED WITH ANOTHER LENS IMMEDIATELY DURING SURGERY; PATIENT HEALTH NOT IMPACTED; PROBLEM CODE: A0414, MATERIAL SPLIT CUT OR TORN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19525243 ↗
  • MalfunctionJul 24, 2023Material Split, Cut or Torn; Material Split, Cut or Torn

    DAMAGED HAPTIC AFTER IMPLANTATION. SURGEON STATED AFTER IMPLANTING THE IOL INTO THE PATIENT'S EYE NOTICED THE FRONT HAPTIC WAS SHEARED OFF WHERE THE BLUE TIP OF THE HAPTIC MEETS CLEAR ACRYLIC. THE LENS WAS LEFT IMPLANTED IN PATIENT'S EYE. SURGEON WILL CONTINUE TO MONITOR FOR ROTATION.

    Patient problems: Insufficient Information

    MAUDE report 17375976 ↗
  • InjuryApr 22, 2022Break; Break

    AFTER LENS WAS IMPLANTED, DOCTOR SAW THE TRAILING HAPTIC WAS BROKEN OFF OF THE LENS. DOCTOR CUT THE LENS TO REMOVED IT WITHOUT INJURY. PATIENT IMPACT: EXPLANTATION.

    Patient problems: Failure of Implant

    MAUDE report 14181462 ↗
  • InjuryMar 23, 2022Sharp Edges; Sharp Edges

    INTRA-OPERATIVE COMPLICATIONS. POSTERIOR CAPSULE RUPTURE. DOCTOR REPORTED THAT SOMETHING HAPPENED IN THE PREPPING OF THE LENS OR IN THE INJECTOR THAT IS CUTTING THE LENS LEADING HAPTIC. THIS IS CREATING A SHARP EDGE THAT IS CAUSING THE POSTERIOR CAPSULE TO TEAR WHEN THE SHARP EDGE ON THE HAPTIC COMES IN CONTACT WITH IT. LENS STILL IN THE EYE VIA REVERSE OPTIC CAPTURE. PATIENT HAS RECOVERED AND IS FINE; PATIENT HAS 2020 AND 2030 UNCORRECTED.

    Patient problems: Capsular Bag Tear

    MAUDE report 13862667 ↗