IPure B1PC (clear), +22.00DRecalls & reported problems

Hoya Medical Singapore Pte. Ltd. · Foldable, posterior chamber intraocular lens (IOL), preloaded in an injector that automatically folds the lens as it is advanced, prior to insertion. Single use. · Class 3 · Implant

All models of IPure (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00886158017240), received between Aug 10, 2021 and Sep 10, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Failure to Fold1
  • Sharp Edges1

Patient problems

  • Capsular Bag Tear3
  • Failure of Implant1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2024Sharp Edges; Sharp Edges

    EVENT OCCURRED IN USA. DAMAGED HAPTIC AFTER IMPLANTATION. WHEN HE WENT TO SPIN THE LENS INTO PLACE IT APPEARED THE HAPTICS TORE THE CAPSULE. ROOT; BROKEN HAPTIC DURING USE; HAPTIC DOES NOT FOLD AS PER IFU. CAPSULAR BAG TEAR / RUPTURE. PRODUCT REMAINED IN EYE AFTER CLINICAL ASSESSMENT. PATIENT HEALTH NOT IMPACTED. PATIENT HAS RECOVERED AND IS FINE. HE LEFT LENS IMPLANTED. BUT QUESTIONED WHETHER THE HAPTICS HAD CHANGED. LENS DEPLOYED AS NORMAL. WHEN HE WENT TO ROTATE THE HAPTICS TORE THE BAG. PROBLEM CODE: A051102 - SHARP EDGES.

    Patient problems: Capsular Bag Tear

    MAUDE report 20185361 ↗
  • InjurySep 12, 2023Failure to Fold; Failure to Fold

    INTRA-OPERATIVE COMPLICATIONS: POSTERIOR CAPSULE RUPTURE, EXPLANTATION. LEAD HAPTIC UNFOLDED INFERIORLY AND BROKE THE BAG AND DOCTOR HAD TO PERFORM A VITRECTOMY. PRODUCT REPLACED WITH ANOTHER LENS IMMEDIATELY DURING SURGERY. PATIENT IMPACT: DEVICE EXPLANTATION.

    Patient problems: Capsular Bag Tear; Failure of Implant

    MAUDE report 17733639 ↗
  • InjuryAug 10, 2021Break; Break

    LEADING HAPTIC BROKE AND BROKE THE BAG POSTERIOR CAPSULE RUPTURE, PRODUCT REPLACED WITH ANOTHER LENS IMMEDIATELY DURING SURGERY; IMPLANTED IN SULCUS; PATIENT HAS RECOVERED AND IS FINE.

    Patient problems: Capsular Bag Tear

    MAUDE report 12304994 ↗