Skip to content

IPure B1PC (clear), +23.00DRecalls & reported problems

Hoya Medical Singapore Pte. Ltd. · Foldable, posterior chamber intraocular lens (IOL), preloaded in an injector that automatically folds the lens as it is advanced, prior to insertion. Single use. · Class 3 · Implant

All models of IPure (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00886158017264), received between Jun 7, 2022 and Sep 10, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Sharp Edges2
  • Failure to Eject1

Patient problems

  • Failure of Implant3
  • Capsular Bag Tear2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2024Sharp Edges; Sharp Edges

    EVENT OCCURRED IN USA. DAMAGED HAPTIC AFTER IMPLANTATION. LEADING; ROOT. WHEN HE WENT TO SPIN THE LENS INTO PLACE IT APPEARED THE HAPTICS TORE THE CAPSULE. BROKEN HAPTIC DURING USE. CAPSULAR BAG TEAR / RUPTURE. PRODUCT REMAINED IN EYE AFTER CLINICAL ASSESSMENT. HE LEFT LENS IMPLANTED; PATIENT HEALTH NOT IMPACTED. PATIENT HAS RECOVERED AND IS FINE; LENS DEPLOYED AS NORMAL. TORIC IOL MISALIGNMENT OR ROTATION. PROBLEM CODE: A051102 - SHARP EDGES

    Patient problems: Capsular Bag Tear

    MAUDE report 20185598 ↗
  • InjuryFeb 16, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    EVENT OCCURRED IN USA. DAMAGED HAPTIC AFTER IMPLANTATION. PROBLEM CODE: A0414, MATERIAL SPLIT CUT OR TORN. TRAILING; LENS WAS IMPLANTED AND THEN HAD TO BE EXPLANTED DUE TO BROKEN TRAIL HAPTIC PATIENT HEALTH NOT IMPACTED; PATIENT HAS RECOVERED AND IS FINE; NO PERMANENT OR NEGATIVE IMPACT ON PATIENT HEALTH EXPECTED. EXPLANTATION DATE: (B)(6) 2023.

    Patient problems: Failure of Implant

    MAUDE report 18724493 ↗
  • InjuryJun 7, 2023Failure to Eject; Failure to Eject

    PROBLEM CODE: FAILURE TO EJECT. CLOGGING OF THE IOL IN THE INJECTOR TIP DURING DELIVERY. IMPLANT DATE: (B)(6) 2023; EXPLANT DATE: (B)(6) 2023 REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 17080647 ↗
  • InjuryFeb 21, 2023Sharp Edges; Sharp Edges

    DR. INSERTED THE LENS INTO THE BAG AND THE BAG TORE. THE DOCTOR FEELS THAT THE LENS HAD SHARP EDGE THAT CAUSED THE BAG TO RIP. LENS WAS EXPLANTED. PATIENT HEALTH WAS NOT IMPACTED.

    Patient problems: Capsular Bag Tear; Failure of Implant

    MAUDE report 16410681 ↗
  • MalfunctionJun 7, 2022Material Split, Cut or Torn; Material Split, Cut or Torn

    DAMAGED HAPTIC AFTER IMPLANTATION. WHILE DOCTOR WAS IMPLANTING THE LENS, HE NOTICED A TEAR AT THE LEADING HAPTIC OPTIC JUNCTION. THE TEAR DID NOT EFFECT THE PATIENT IN THIS SURGERY. PATIENT HEALTH NOT IMPACTED; PATIENT HAS RECOVERED AND IS FINE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14621694 ↗

Other Intraocular Lens devices

All Intraocular Lens devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page