SKATER™ Drainage Catheter 755108035ERecalls & reported problems

Argon Medical Devices, Inc. · SKATER™ Drainage Catheter - locking pigtail 8F x 35cm · Class 1

All models of SKATER Drainage Catheter (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00886333000166), received between May 4, 2023 and Nov 22, 2024 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak17

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Insufficient Information5

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 22, 2024Fluid/Blood Leak; Fluid/Blood Leak

    DURING THE PERCUTANEOUS NEPHROSTOMY PROCEDURE, THE DOCTOR FOUND THAT THE PIGTAIL'S SUTURE SLEEVE AND SUTURE STRINGS WERE LEAKING BODILY FLUIDS. THIS LEAKAGE ISSUE MADE THE DOCTOR VERY ANGRY BECAUSE IT KEPT OCCURRING CONTINUOUSLY, FORCING THE DOCTOR TO REPLACE THE PIGTAIL TO SOLVE THE PROBLEM. ADDITIONALLY, IT CAUSED FURTHER HARM TO THE PATIENT. OUR CUSTOMERS BELIEVE THAT ARGON MEDICAL DOES NOT VALUE THIS FEEDBACK, WHICH IS WHY THIS POOR EXPERIENCE OCCURS REPEATEDLY EVERY MONTH. THE FAULTY PRODUCT RETURNED FROM THE HOSPITAL INCLUDED ONLY THE PIGTAIL, WHICH WAS FOUND TO BE CUT INTO TWO PIECES. A VIDEO WAS PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 20762452 ↗
  • MalfunctionApr 11, 2024Fluid/Blood Leak; Fluid/Blood Leak

    A REPORTER INDICATED THERE WAS LEAKAGE FROM THE HUB AFTER FINISHED CASE. THE TREATMENT PROCESS WAS COMPLETED, BUT THE AREA WHERE THE THREAD CAME OUT LEAKED, SO HAD TO RE-TREAT IT WITH A DIFFERENT PRODUCT.

    Patient problems: Insufficient Information

    MAUDE report 19087657 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17052604 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17052475 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17052297 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17052130 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17051135 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIO-CONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17051030 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050842 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050668 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050505 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050369 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17050241 ↗
  • MalfunctionJun 2, 2023Fluid/Blood Leak; Fluid/Blood Leak

    PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17049872 ↗
  • MalfunctionMay 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SAME PROBLEM IS OCCURRING IN SEVERAL PATIENTS (12 AT LEAST): PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES; ALTHOUGH THE HOSPITAL NEEDS FREE UNITS OF DEVICE FOR REPLACEMENTS FROM DIFFERENT LOTS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16871172 ↗
  • MalfunctionMay 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SAME PROBLEM IS OCCURRING IN SEVERAL PATIENTS (12 AT LEAST): PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES; ALTHOUGH THE HOSPITAL NEEDS FREE UNITS OF DEVICE FOR REPLACEMENTS FROM DIFFERENT LOTS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16870984 ↗
  • MalfunctionMay 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    SAME PROBLEM IS OCCURRING IN SEVERAL PATIENTS (12 AT LEAST): PHYSICIAN REPORTS THAT IN THE CONNECTOR OF THE IMPLANTED CATHETER, THROUGH THE HOLE WHERE ITS ANCHOR THREAD PASSES, THE LIQUID THAT THE CATHETER IS DRAINING ALSO COMES OUT; BECAUSE THAT OUTLET HOLE IS NOT WELL SEALED AND THE INTERNAL VALVE OF THE CATHETER DOES NOT WORK EITHER. THE RESULT IS THAT THE DOCTOR MUST CHANGE THE CATHETER AND IS A DANGER BECAUSE OF THE GREAT POSSIBILITY OF INFECTIONS OF THE AREA WHERE THE CATHETER COMES OUT. OBTAINING SAMPLES FROM THE FAILED CATHETERS EXPLANTED IS PRACTICALLY IMPOSSIBLE DUE TO THE GREAT POSSIBILITY OF DANGEROUS BIOCONTAMINATION DURING THE HANDLING OF THE RESCUED DEVICES; ALTHOUGH THE HOSPITAL NEEDS FREE UNITS OF DEVICE FOR REPLACEMENTS FROM DIFFERENT LOTS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16869861 ↗

Other Catheter, Continuous Irrigation devices

All Catheter, Continuous Irrigation devices →