Skip to content

Cohere 1814365Recalls & reported problems

Nuvasive, Inc. · COHERE TLIF-A Inserter, Offset · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517017581), received between Oct 10, 2023 and Oct 10, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Activation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Problem1
  • Device Damaged by Another Device1
  • Mechanical Problem1
  • Positioning Problem1
  • Unintended Movement1

Patient problems

  • Unspecified Nervous System Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 10, 2023Mechanical Problem; Device Damaged by Another Device; Positioning Problem; Unintended Movement; Activation Problem; Mechanical Problem; Device Damaged by Another Device; Positioning Problem; Unintended Movement; Activation Problem

    REP REPORTED THAT DURING A TRANSFORAMINAL LUMBAR INTERBODY FUSION PROCEDURE THE RESIDENT PLACED THE CAGE WITHOUT ISSUE BUT WAS UNCOMFORTABLE WITH THE POSITION, SO THE INSERTER WAS REATTACHED, AND SLAP HAMMERED BACK OUT. DURING THE STATED REMOVAL THE INSERTERS (1814364 LOT TI1718 STRAIGHT) DOWEL PIN CAME LOOSE, AND IT WAS NO LONGER USABLE. THE CAGE WAS THEN ATTACHED TO THE OFF-SET INSERTER (1814365 LOT TI1580) FOR PLACEMENT. DURING THIS SECOND IMPACTION OF THE SAME CAGE IN-CASE NEURAL MONITORING OF THE PATIENT (THIRD PARTY CADWELL) IDENTIFIED MOTOR DEFICIT HAD TAKEN PLACE AND THE APPROACH WAS ABORTED, THE RESIDENT CLAIMS THAT THE INSERTER WAS ALLOWING THE IMPLANT TO PIVOT PREMATURELY IT BECAME LOOSE AND DID NOT ENGAGE THE INTERBODY FULLY DURING IMPLANTATION. THE PEEK INTERBODY WAS DISCARDED AT THE USER FACILITY AND UNAVAILABLE FOR EVALUATION.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 17906238 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →