Helix 7741115Recalls & reported problems
Nuvasive, Inc. · Helix Mini Screw, 4x15mm Var Self-Drill · Class 2 · Implant
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517096784), received between Sep 23, 2021 and Mar 28, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device2
- Device Damaged by Another Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 28, 2024Device Damaged by Another Device; Device Damaged by Another Device
ON (B)(6) 2021 A PATIENT UNDERWENT A C5/6 ANTERIOR CERVICAL DISCECTOMY, BILATERAL C6 FORAMINOTOMY, AND FUSION AND PLATING PROCEDURE, UTILIZING A MODULUS-C INTERBODY AND A HELIX MINI PLATE WITH HELIX MINI SCREWS, TO ADDRESS BILATERAL C6 RADICULOPATHY, THERE WERE NO NOTED COMPLICATIONS DURING THIS PROCEDURE. IT WAS REPORTED THE PATIENT SUBSEQUENTLY EXPERIENCED ONGOING PAIN WITH SUSPECTED NON-UNION AND IT WAS LATER DISCOVERED THAT ONE OF THE SCREWS HAD LOST FIXATION. ON (B)(6) 2022 A REVISION SURGERY WAS PERFORMED WHERE THE PLATE, SCREWS, AND INTERBODY WERE REMOVED. AFTER REMOVAL, IT WAS NOTED ONE OF THE SCREWS WAS MISSING ITS COLLAR COMPONENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Pain
MAUDE report 18996259 ↗MalfunctionSep 23, 2021Malposition of Device; Malposition of Device
ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE ON UNKNOWN LEVELS OF THE SPINE. ON (B)(6) 2021 IT WAS DISCOVERED THAT DURING A ROUTINE FOLLOW UP THAT AN IMPLANTED SCREW ADVANCED THROUGH THE CANTED COIL LOCKING MECHANISM OF AN IMPLANTED PLATE AND INTO THE VERTEBRAL BODY. THE PATIENT IN ASYMPTOMATIC AND NO REVISION IS PLANNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12522561 ↗MalfunctionSep 23, 2021Malposition of Device; Malposition of Device
ON (B)(6) 2021 A PATIENT UNDERWENT A AN ANTERIOR CERVICAL DISCECTOMY PROCEDURE ON UNKNOWN LEVELS OF THE SPINE. ON A UNKNOWN DATE IT WAS DISCOVERED DURING A ROUTINE FOLLOW UP THAT A IMPLANTED SCREW ADVANCED THROUGH THE CANTED COIL LOCKING MECHANISM OF AN IMPLANTED PLATE AND INTO THE VERTEBRAL BODY. THE PATIENT IS ASYMPTOMATIC AND NO ADVERSE CONSEQUENCES WERE REPORTED. A REVISION OCCURRED ON SEPTEMBER (B)(6) 2021 WHERE THE DEVICE WAS REMOVED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12522367 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.