NVM5 2021000Recalls & reported problems
Nuvasive, Inc. · NVM5 Control Unit-2 · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517168603), received between Aug 21, 2022 and Aug 23, 2023 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
1 other recall cover devices cleared under the same FDA submission (K143641); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem1
- Electromagnetic Interference1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Insufficient Device Problem Information1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Nerve Damage1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryAug 23, 2023Communication or Transmission Problem; Insufficient Device Problem Information; Communication or Transmission Problem; Insufficient Device Problem Information
MalfunctionAug 29, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
DURING SERVICING OF THE NVM5 UNIT, IT WAS FOUND THE UNIT FAILED SIMULATED STIMULATION TESTING; SPECIFICALLY, THE UNIT WAS NOT PROVIDING APPROPRIATE RESPONSES TO RETURNED STIMULATION VALUES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15317126 ↗MalfunctionAug 21, 2022Electromagnetic Interference; Electromagnetic Interference
ON (B)(6) 2022 DURING AN EXTREME LATERAL INTERBODY FUSION AN NVM5 SYSTEM DISPLAYED A NOTIFICATION THAT INDICATED THERE WAS "ELECTOCAUTERY INTERFERENCE" AND STIMULATION COULD NOT BE PERFORMED. TROUBLESHOOTING WAS PERFORMED BY SWITCHING OFF THE BED AND ELECTOCAUTERY UNIT, ENSURING THE X-RAY WAS AT A DISTANCE, AND REPLACING THE GROUND ELECTRODES; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. THE SURGEON ALTERED THE SURGICAL PLAN AND PROCEEDED WITH A POSTERIOR TRANSFORAMINAL LUMBAR INTERBODY FUSION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15263566 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143641. One submission often covers many models, so check the product and lot details in each record.
- Z-2684-2020 · Jun 10, 2020 · TerminatedNuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal s
Other STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls795 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls331 reports
- NIM Vital NIM4CPB1Medtronic Xomed, Inc. · Stimulator, electrical, evoked response2 recalls208 reports
- POWEREASE™ Instruments 2300000Medtronic Sofamor Danek, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE1 recall203 reports
- NIM Vital™ NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE3 recalls47 reports
- NIM-Response® 8253200Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE113 reports
- NA DAQ916Medtronic Xomed, Inc. · STIMULATOR, ELECTRICAL, EVOKED RESPONSE80 reports
- Neuromaster G1 DC-200BNihon Kohden Corporation · STIMULATOR, ELECTRICAL, EVOKED RESPONSE73 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page