Skip to content

NVM5 2029950Recalls & reported problems

Nuvasive, Inc. · NVM5 XLIF Dilator Kit, Pkg Ster · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517170606), received between Jun 1, 2018 and Feb 2, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material1
  • Imprecise or Erratic Results1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 2, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING NEUROMONITORING VIA ELECTROMYOGRAPHY IN AN EXTREME LATERAL INTERBODY FUSION PROCEDURE, INCONSISTENT VALUES RELATED TO NERVE PROXIMITY/DIRECTIONALITY WERE OBTAINED WHILE USING THE DILATORS. SPECIFICALLY, THE 6MM DILATOR GAVE VALUES OF =20MA WHILE THE 9MM AND 12MM DILATORS GAVE VALUES OF <5MA IN ALL DIRECTIONS. ATTEMPTS TO RESOLVE BY REPOSITIONING THE DILATORS WERE UNSUCCESSFUL. REPORTEDLY, SMALL PARTICLES WERE NOTED TO BE ATTACHED TO THE K-WIRE AFTER IT WAS REMOVED FROM THE 6MM DILATOR. A NEW SET OF DILATORS WAS AVAILABLE AND USED TO SUCCESSFULLY COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18634416 ↗
  • MalfunctionOct 25, 2022Imprecise or Erratic Results; Imprecise or Erratic Results

    THIS IS A DUPLICATE OF REPORT 2031966 - 2022 - 00223 AND FOLLOW UP REPORT 2031966 - 2022 - 00223 DETAILS ON SEC H10.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15670610 ↗
  • MalfunctionJun 1, 2018Leak/Splash; Device Contamination with Chemical or Other Material

    LATERAL LUMBAR INTERBODY FUSION (XLIF) WITH DR. NUVASIVE NVM5 XLIF DISPOSABLE KIT, REF: 2029950, LOT: 1802231, EXP: 2023-01-22, HAD BLACK INK LIKE SUBSTANCE COME OFF OF DILATOR (LOOKS LIKE FROM THE INSIDE OF THE BLUE (LARGEST) DILATOR). CONTAMINATED ITEMS REMOVED FROM SURGICAL FIELD. DILATORS WITH PACKAGING SENT TO RISK MANAGEMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7560685 ↗

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →