Skip to content

Armada 8448590Recalls & reported problems

Nuvasive, Inc. · ARM15T Screw, 8.5x90mm Iliac SL-Open · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517256492), received between Jan 5, 2017 and Jan 5, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 12/12/2016 IT WAS REPORTED THAT DURING A THORACIC LUMBAR SURGERY ON (B)(6) 2016, THE TULIP HEAD OF ONE ILIAC SCREW DISENGAGED AND THE THREAD IN THE SHANK WAS STRIPPED. THE SCREW WAS LEFT IN THE PATIENT AND NO PATIENT INJURY WAS REPORTED. THE SURGEON HAS ELECTED TO MONITOR THE PATIENT. NO REVISION SURGERY IS SCHEDULED AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 6226208 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page