Skip to content

X-CORE 7180027ARecalls & reported problems

Nuvasive, Inc. · X-CORE 2 Ti Core, Ø18x21-27mm Autolock · Class 2 · Implant

All models of X-CORE (29)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517365811), received between Sep 8, 2025 and Sep 8, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Jam” (1 report).

1 other recall cover devices cleared under the same FDA submission (K090176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 8, 2025Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE X-CORE FAILED TO EXPAND DURING A CASE AND ANOTHER IMPLANT WAS USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22998433 ↗

FDA clearance and classification

Cleared through Special 510(k) K090176 on Apr 27, 2009 (decision: Substantially Equivalent). Submitted by Nuvasive, Inc. as “Nuvasive Adjustable Vertebral Body Replacement System”. The FDA review took 94 days. FDA record ↗

The FDA regulates product code MQP (spinal vertebral body replacement device) under 21 CFR 888.3060 as a Class II device. 21 CFR 888.3060 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K090176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0738-2023 · Nov 10, 2022 · Open, Classified
    X-CORE 2 Ti Core, Static. Part of the NuVasive X-Core Expandable VBR System used with spinal fixation.

Common questions about the X-CORE 7180027A

Has the X-CORE 7180027A been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887517365811), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the X-CORE 7180027A?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 8, 2025. That report lists device problem “Mechanical Jam” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the X-CORE 7180027A?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the X-CORE 7180027A FDA approved?

It was cleared through Special 510(k) K090176 on Apr 27, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the X-CORE 7180027A MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the X-CORE 7180027A, and is it still sold?

The FDA's device database lists Nuvasive, Inc. as the company responsible for it (brand X-CORE, model 7180027A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices

All SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page