Skip to content

Precept 8896540Recalls & reported problems

Nuvasive, Inc. · PRECEPT Screw Shank, 6.5x40mm Modular · Class 2 · Implant

All models of Precept (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517387530), received between Sep 12, 2017 and Sep 12, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Failure of Implant1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 12, 2017Fracture; Fracture

    IT WAS REPORTED DURING A ROUTINE FOLLOW UP VISIT IT WAS DISCOVERED VIA X-RAY THAT THE SCREW ON THE LEFT SIDE OF THE S1 LEVEL CONSTRUCT HAD FRACTURED. PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2017 WHERE THE BROKEN SCREWS AND RODS WERE REPLACED AND AN ADJACENT LEVEL WAS ADDED TO THE CONSTRUCT AT THE L3 LEVEL. DURING THE PROCEDURE IT WAS OBSERVED THAT THE S1 SCREW HAD FRACTURED ON THE RIGHT SIDE OF THE CONSTRUCT. WHEN ATTEMPTING TO REMOVE THE SCREWS HALF OF THE FRACTURED SCREW COULD NOT BE RETRIEVED FROM THE PATIENT AND WAS LEFT IN BONE. THE SURGEON REPORTED PSEUDARTHROSIS WAS OBSERVED AND BELIEVED MAY HAVE CONTRIBUTED TO THE EVENT. NO ADVERSE ISSUES TO THE PATIENT WERE REPORTED. IT WAS REPORTED THE PATIENT IS DOING WELL POST-REVISION SURGERY.

    Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture

    MAUDE report 6859696 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →