Skip to content

X-CORE 7160025Recalls & reported problems

Nuvasive, Inc. · X-CORE 2 Ti Core, Ø16x18-25mm · Class 2 · Implant

All models of X-CORE (29)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517424235), received between Sep 1, 2021 and Dec 31, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

1 other recall cover devices cleared under the same FDA submission (K090176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Advance1
  • Migration1

Patient problems

  • Failure of Implant1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT OPERATION ON (B)(6) 2025; FIXATION LEVEL: TH10 ¿ L3: PS, L1 :X-CORE, L2/3 PLIF. THE CONDITION WAS AN L1 BURST FRACTURE. THE SURGEON INSERTED COROENT LM INTO THE L2/L3 INTERVERTEBRAL SPACE BY PLIF AND PERFORMED POSTERIOR FIXATION FROM TH10 TO L3. AND THEN, THE SURGEON THEN CHANGED THE PATIENT'S POSITION AND REPLACED THE L1 TO X-CORE VIA AN ANTERIOR APPROACH. DURING THE SURGERY, THERE WAS MINIMAL BLEEDING AND THE PATIENT'S VITAL SIGNS WERE STABLE. HOWEVER, THE PATIENT BUT DID NOT REGAIN CONSCIOUSNESS AND EXHIBITED OCCASIONAL CONVULSIONS. CT SCANS WERE PERFORMED FROM THE HEAD (INCLUDING THE IMPLANT SITE) TO THE ABDOMEN, BUT THE SURGEON COULD NOT IDENTIFY THE CAUSE. THE PATIENT TREATED IN THE ICU WHILE THE CAUSE IS BEING INVESTIGATED. ON (B)(6), THE PATIENT HAD AN MRI SCAN TO FIND THE CAUSE, AND THERE WERE NO FINDINGS IN THE BLOOD VESSELS AND IMPLANTS. WE DID NOT OBTAIN ANY COMMENTS REGARDING A CAUSAL RELATIONSHIP WITH OUR IMPLANTS AND MEDICAL DEVICES FROM THE SURGEON. THE SURGEON SUSPECTED NEUROLOGICAL DAMAGE DUE TO AN EPILEPTIC SEIZURE, CEREBRAL INFARCTION, OR PULMONARY EMBOLISM, BUT COULD NOT IDENTIFY THE CAUSE. ON (B)(6): THE PATIENT SHOWE

    Patient problems: Insufficient Information

    MAUDE report 23933826 ↗
  • MalfunctionNov 3, 2025Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT DURING SURGERY THE X-CORE 2 FAILED TO EXPAND. THE SURGEON REMOVED IT AND IMPLANTED ANOTHER ONE. THIS EVENT OCCURRED IN ITALY.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23457944 ↗
  • MalfunctionSep 1, 2021Migration; Migration

    ON (B)(6) 2021 A PATIENT UNDERWENT A CORPECTOMY THE FIRST OF A TWO-PART PROCEDURE AT T12. ON THE SECOND PART POSTERIOR FIXATION AT T10/L2 WAS COMPLETED. ON (B)(6) 2021 THE PATIENT REPORTED PAIN AND RADIOGRAPHS REVEALED SUBSIDENCE AT T12 AND FOUR SCREWS AT L1 AND L2 BACKED OUT. ON (B)(6) 2021 A REVISION SURGERY WAS CONDUCTED WHERE THE FOUR SCREWS AT L1/2 WERE REPLACED AND THE CONSTRUCT WAS EXTENDED TO T10/L4.

    Patient problems: Pain

    MAUDE report 12411339 ↗

FDA clearance and classification

Cleared through Special 510(k) K090176 on Apr 27, 2009 (decision: Substantially Equivalent). Submitted by Nuvasive, Inc. as “Nuvasive Adjustable Vertebral Body Replacement System”. The FDA review took 94 days. FDA record ↗

The FDA regulates product code MQP (spinal vertebral body replacement device) under 21 CFR 888.3060 as a Class II device. 21 CFR 888.3060 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K090176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0738-2023 · Nov 10, 2022 · Open, Classified
    X-CORE 2 Ti Core, Static. Part of the NuVasive X-Core Expandable VBR System used with spinal fixation.

Common questions about the X-CORE 7160025

Has the X-CORE 7160025 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887517424235), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the X-CORE 7160025?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Sep 1, 2021 and Dec 31, 2025. In those reports, the most common specific device problems are “Failure to Advance” (1) and “Migration” (1); the most common patient problems are “Failure of Implant” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the X-CORE 7160025?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the X-CORE 7160025 FDA approved?

It was cleared through Special 510(k) K090176 on Apr 27, 2009. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the X-CORE 7160025 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the X-CORE 7160025, and is it still sold?

The FDA's device database lists Nuvasive, Inc. as the company responsible for it (brand X-CORE, model 7160025). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices

All SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page