Skip to content

ReLine 12355300Recalls & reported problems

Nuvasive, Inc. · RELINE MAS Ti Rod, 5.5x300mm Straight · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887517506764), received between Jun 2, 2017 and Nov 3, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2021Fracture; Fracture

    ON (B)(6) 2021 A PATIENT UNDERWENT POSTERIOR LUMBAR INTERBODY FUSION PROCEDURE ON UNKNOWN LEVELS OF THE SPINE AND THE SURGERY WAS COMPLETED WITHOUT ISSUE. ON A UNKNOWN DATE NEAR THE END OF (B)(6) THE PATIENT WAS BENDING OVER AND FELT A POP. ON (B)(6) THE PATIENT FOLLOWED UP WITH THE SURGEON AND IT WAS DISCOVERED VIA RADIOGRAPH THAT A IMPLANTED ROD HAD FRACTURED. A REVISION OCCURRED ON (B)(6) 2021 WHERE THE FRACTURED ROD WAS REPLACED. DURING THE REVISION THE DOCTOR NOTED PSEUDO ARTHROSIS AT L4-5, L5-S1. THE PATIENT IS DOING WELL POST REVISION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12747778 ↗
  • MalfunctionJun 2, 2017Fracture; Fracture

    ON (B)(6) 2016 A PATIENT UNDERWENT A MULTI-LEVEL INTERBODY PLACEMENT PROCEDURE WITH POSTERIOR FIXATION FROM T9 TO S1 WITHOUT ISSUE. APPROXIMATELY ONE MONTH LATER DURING FOLLOW UP A ROD FRACTURE WAS OBSERVED ON RADIOGRAPHS. NO REVISION IS PLANNED AT THIS TIME. NO PATIENT INJURY HAS BEEN REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 6608773 ↗

Other Pedicle screw spinal system, adolescent idiopathic scoliosis devices

All Pedicle screw spinal system, adolescent idiopathic scoliosis devices →