Skip to content

ReLine 13116545Recalls & reported problems

Nuvasive, Inc. · RELINE-O Screw, 6.5x45mm Traction Poly · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00887517582508), received between Apr 11, 2018 and Dec 4, 2023 — classified as 3 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Fracture1
  • Malposition of Device1
  • Migration or Expulsion of Device1

Patient problems

  • Bone Fracture(s)2
  • Failure of Implant2
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1
  • Paralysis1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(6) 2023:UNDERWENT PLF (POSTEROLATERAL FUSION) BY RELINE WAS PERFORMED AT TH10/L5. (B)(6) 2023: A VERTEBRAL FRACTURE OCCURRED AT TH9 DUE TO ADJACENT SEGMENT DISEASE AND THE PATIENT HAD MILD PARALYSIS FROM UNIDENTIFIED SOURCE, SO THE FIXATION LEVEL WAS EXTENDED TO TH6.

    Patient problems: Bone Fracture(s); Paralysis

    MAUDE report 18257057 ↗
  • MalfunctionAug 21, 2022Fracture; Fracture

    IT WAS REPORTED THE PATIENT UNDERWENT AN INITIAL POSTERIOR FIXATION PROCEDURE. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION PROCEDURE APPROXIMATELY TWO (2) YEARS LATER TO REVISE THE FIXATION CONSTRUCT TO ADDRESS SYMPTOMS AT ADJACENT LEVELS. DURING THE REVISION, IT WAS IDENTIFIED THAT ONE OF THE SCREWS AT THE L3 LEVEL HAD FRACTURED AS ONLY THE PROXIMAL PORTION OF THE SCREW SHANK WAS ABLE TO BE REMOVED. THE DISTAL END OF THE SCREW SHANK WAS UNABLE TO BE REMOVED FROM THE VERTEBRAL BODY BUT WAS REPORTED TO BE COMPLETELY ENCASED IN THE BONE. AS A RESULT, THE FRACTURED PORTION WAS RETAINED BY THE PATIENT AND THE FIXATION CONSTRUCT WAS EXTENDED TO THE LEVEL BELOW. NO FURTHER PATIENT IMPACT WAS REPORTED AND THE PATIENT REPORTEDLY DID NOT EXPERIENCE PAIN OR ANY SYMPTOMS RELATED TO THE FRACTURED SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15263577 ↗
  • MalfunctionFeb 4, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT L2/4. ON AN UNKNOWN DATE IT WAS DISCOVERED THAT THE INTERBODY WAS DISPLACED LATERALLY (200061169) DUE TO THE LOOSENING OF A SPINAL SCREW AT L2 OR L4. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED WHERE THE L2 AND L4 SCREWS WERE REPLACED AND COMPETITOR'S INTERBODIES WERE INSERTED AT L4/5/S. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13455713 ↗
  • MalfunctionJun 10, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    A REVISION PROCEDURE WAS CONDUCTED ON (B)(6) 2019 DUE TO ADJACENT SEGMENT DISEASE (ASD). AS PER REPORTER THE IMPLANTED SCREWS AT L4 LEVEL HAD BECAME LOOSEN, THIS WAS NOT THE CAUSE OF THE REVISION PROCEDURE BUT, RATHER WAS DISCOVERED DURING THE REVISION. THE PHYSICIAN REPLACED THESE SCREWS WITH NEW ONES AT L4 AND ADDED NEW SCREW & OFFSET CONNECTOR AT ILIUM.

    Patient problems: Failure of Implant

    MAUDE report 10139337 ↗
  • InjuryAug 21, 2018Malposition of Device; Malposition of Device

    ON (B)(6) 2018, PATIENT UNDERWENT AN EXTREME LATERAL INTERBODY FUSION PROCEDURE AND POSTERIOR FIXATION ON (B)(6) 2018 AT LEVELS L2-L5 WITH NO REPORTED ISSUES. POST-OPERATIVE EXAMINATION FOUND THE RIGHT SIDE SCREW AT THE L5 LEVEL WAS PLACE INSIDE-OUT INSTEAD OF THE INTENDED OUTSIDE-IN. ON (B)(6) 2018 A REVISION PROCEDURE WAS PERFORMED REPLACING AND REDIRECTING THE SCREW. NO ALLEGED PRODUCT MALFUNCTION REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7803931 ↗
  • InjuryJun 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2018 PATIENT UNDERWENT A PROCEDURE TO REVISE POSITIONING OF THE LEFT L3, L4 AND ILIAC SCREWS. THE SCREWS AT L2 LEVEL WERE FOUND TO BE LOOSE AND WERE REPLACED, THE SCREWS REMOVED AT L2 WERE REUSED AND PLACED AT L1. ON (B)(6) 2018 IT WAS DISCOVERED THAT THE L1 VERTEBRAL BODY WAS FRACTURED AND A REVISION WAS PERFORMED TO EXTEND THE CONSTRUCT FROM L1 TO T9 LEVEL.

    Patient problems: Bone Fracture(s)

    MAUDE report 7623255 ↗
  • MalfunctionApr 11, 2018Migration or Expulsion of Device; Migration or Expulsion of Device

    ON (B)(6) 2018, A PATIENT UNDERWENT AN EXTREME LATERAL INTERBODY FUSION PROCEDURE AT L2-L5 AND POSTERIOR FUSION PROCEDURE AT L2-ILIAC LEVELS. REPORTEDLY, ON (B)(6) 2018, IT WAS CONFIRMED THAT THE LEFT L3 AND L4 SCREWS HAD PENETRATED THE INNER BONE AND THE LEFT ILIAC SCREW HAD ALSO PENETRATED THE OUTER TABLE OF THE ILIUM. ON (B)(6) 2018 PATIENT UNDERWENT A REVISION PROCEDURE WHERE THE LEFT L3, L4 AND ILIAC SCREWS WERE REPLACED. SCREWS ON THE L2 LEVEL WERE FOUND TO BE LOOSE AND WERE REPLACED WITH 7.5MM SCREWS, THE SCREWS REMOVED WERE PLACED AT THE L1 LEVEL.

    Patient problems: Failure of Implant

    MAUDE report 7423427 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates