ICU Medical B2020Recalls & reported problems
Icu Medical, Inc. · 60" (152 cm) Appx 0.4mL, Smallbore Ext Set w/Clamp, Luer Lock · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00887709002937), received between Dec 27, 2024 and Aug 5, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
188 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- Crack5
- No Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
THE COMPLAINT INVOLVED A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK DEVICE. A SMALL COLLECTION OF FLUID WAS REPORTED ON THE SYRINGE PUMP BELOW THE CONNECTION FROM THE SYRINGE TO THE TUBING. THE PUMP WAS NOTED TO FEEL WET ON THE BOTTOM. ALSO, THEY NOTED A SMALL COLLECTION OF FLUID ON THE FLOOR BELOW THE IV POLE. THE STAFF ASSUMED IT WAS CAUSED FROM A LOOSE CONNECTION; AND THE TUBING WAS CHANGED WITH A NEW SYRINGE. THE STAFF TESTED THE CONNECTION BY FLUSHING THE PORT WITH SALINE AND NOTED TO BE A CRACK/LEAK NOTED ON THE IV TUBING BELOW THE CLAVE. THERE WAS UNEXPECTED OR PROLONGED CARE. THE REPORT STATED THAT THE STAFF WENT INTO ROOM TO CHANGE THE EMPTY DEXMEDETOMIDINE SYRINGE. THE TUBING WAS ATTACHED TO A MICROCLAVE AND 10 ML PREFILLED SALINE SYRINGE THESE ITEMS WILL BE RETURNED FOR EVALUATION. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22719522 ↗MalfunctionJul 14, 2025Fluid/Blood Leak; Fluid/Blood Leak
A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK THAT WAS REPORTEDLY LEAKING DURING USE ON A PATIENT, ADMINISTERING BIVALRUDIN. THERE WAS NO ADVERSE EVENT NOR ANY DELAY IN THERAPY THAT WOULD LEAD TO PATIENT HARM IF NOT CAUGHT BY STAFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22502864 ↗MalfunctionJun 26, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED THAT A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK WAS FOUND TO HAVE EXPERIENCED A LEAK DURING PATIENT USE. IT WAS REPORTED THAT THAT FLUID WAS DRIPPING TO THE FLOOR BY THE INTRAVENOUS (IV) POLE. UPON INVESTIGATION, MORPHINE WAS FOUND TO BE LEAKING FROM THE TUBING WHERE THE SMARTSITE CONNECTS TO THE TUBING. ADDITIONALLY, THERE WAS A CRACK IN THE TUBING. IT WAS ALSO MENTIONED THAT THE PATIENT HAD ABNORMAL VITALS (TEMPERATURE/HR, RR), WORK OF BREATHING, AND AGITATION EMESIS. NO ADDITIONAL INFORMATION WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22341202 ↗MalfunctionMay 16, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
THE EVENT INVOLVED A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK WHERE IT WAS REPORTED THAT THERE IS A CRACK IN THE TUBING CAUSING IT TO LEAK WHEN ADMINISTERING MORPHINE AT THE SMARTSITE CONNECTION. THERE WAS PATIENT INVOLVEMENT AND NO HUMAN HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22036181 ↗MalfunctionMay 16, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
THE EVENT INVOLVED A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK WHERE IT WAS REPORTED THAT THERE IS A CRACK IN THE TUBING CAUSING IT TO LEAK WHEN ADMINISTERING CEFEPIME AT THE SMARTSITE CONNECTION. THERE WAS PATIENT INVOLVEMENT AND NO HUMAN HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22035946 ↗MalfunctionMay 8, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
A COMPLAINT WAS RECEIVED REGARDING A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK THAT EXPERIENCED LEAKAGE DURING PATIENT USE. IT WAS REPORTED THAT THERE IS A CRACK IN THE TUBING CAUSING IT TO LEAK WHEN ADMINISTERING MEDICATION AT THE SMARTSITE CONNECTION. THE MEDICATION THAT WAS LEAKING AT THE TIME OF THE EVENT WAS SEDATION DRUG. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO ADVERSE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21989265 ↗MalfunctionDec 27, 2024No Flow; No Flow
A COMPLAINT WAS RECEIVED REGARDING A 60" (152 CM) APPX 0.4ML, SMALLBORE EXT SET W/CLAMP, LUER LOCK THAT EXPERIENCED NO FLOW. IT WAS REPORTED THAT THE "PUMP ERROR CODE INDICATING 'HIGH PRESSURE' IN THE SYRINGE PUMP, RESISTANCE. IMPOSSIBLE TO ADMINISTER MANNITOL WITH THE SYRINGE PUMP, LOSS OF TIME AND INCREASED WORKLOAD FOR THE NURSE. DELAY IN ADMINISTERING MEDICATION TO THE USER. PRODUCTS THAT CAUSED ERROR CODES HAVE BEEN DISCARDED. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21031128 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.
- Z-3163-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-3157-2026 · May 12, 2026 · Open, ClassifiedIV Non-Dedicated Gravity Administration Sets with drip chambers
- Z-0154-2026 · Oct 2, 2025 · Open, ClassifiedIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
- Z-1721-2022 · Aug 1, 2022 · Open, Classified6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
- Z-0957-2021 · Dec 2, 2020 · Terminated7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
- Z-0958-2021 · Dec 2, 2020 · Terminated154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
- Z-0956-2021 · Dec 2, 2020 · Terminated60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
- Z-0955-2021 · Dec 2, 2020 · Terminated60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
- Z-0214-2021 · Jul 6, 2020 · Terminated99 IN (251cm) APPX 12.6ml, 20 Drop, 2 Clave, 2 NanoClave 4-Way Stopcocks, Drop-In Ext, Clave; 1 unit per pouch, 25 pouches per case. UDI: (01)10887709066318(17)241201(30)25(10)4515
- Z-0255-2021 · Jul 6, 2020 · Terminated148 IN (376cm) APPX 18.8 ml, 15 DROP SET, 2 CLAVE, Remv 3 Gang Clave Stopcock.1 unit per pouch, 25 pouches per case. UDI:(01)10887709071558(17)250101(30)25(10)4603841. The Intrav
- Z-0199-2021 · Jul 6, 2020 · Terminated9.5 IN (24cm)APPX 1.1ml SMALLBORE TRIFUSE SET,3 MicroClave CLEAR, NanoClave, 3 CHECK VALVES; 1 unit per pouch, 50 pouches per case. UDI:(01)10887709069340(17)250101(30)50(10)455847
- Z-0245-2021 · Jul 6, 2020 · Terminated126 IN (320cm) APPX 23.5ml, 10 Drop Set 3 Clave, Remv Clave Stopcock, 2 Ext.1 unit per pouch, 25 pouches per case. UDI:(01)10887709062365(17)250201(30)25(10)4630752. The Intrav
- Z-0233-2021 · Jul 6, 2020 · Terminated139" (353 cm) 15 Drop Admin Set w/3 MicroClave Clear, 4-Way Stopcock, 3-Port NanoClave Manifold w/Check Valve, Spin Luer, 2 Ext. 1 unit per pouch, 25 pouches per case. UDI:(01)1088
- Z-0195-2021 · Jul 6, 2020 · Terminated109" (277 cm) Appx 8.6 ml, Transfer Set w/Check Valve, NanoClave T-Connector, Anti-Siphon Valve, 4 Clamps, Smallbore Trifuse Ext Set w/2 MicroClave Clear, 0.2 Micron Filter, Rotati
- Z-0258-2021 · Jul 6, 2020 · Terminated112" (284 cm) 60 Drop 150ml Burette Set (Clave, Shut-Off) w/2 MicroClave, Remv 3 Gang 4-Way Clave Stopcock, Check Valve, Rotating Luer, 1 Ext. 1 unit per pouch, 25 pouches per case
…and 173 more.
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page