Skip to content

ICU Medical B9319Recalls & reported problems

Icu Medical, Inc. · 113" (287 cm) Appx 14.4 mL, 10 Drop Primary Set w/2 MicroClave®, Rotating Luer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

187 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 12, 2026Class IIOpen, Classified

    Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

    Z-3163-2026 · Root cause: Process control · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.

  • Z-3157-2026 · May 12, 2026 · Open, Classified
    IV Non-Dedicated Gravity Administration Sets with drip chambers
  • Z-0154-2026 · Oct 2, 2025 · Open, Classified
    IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
  • Z-1721-2022 · Aug 1, 2022 · Open, Classified
    6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
  • Z-0957-2021 · Dec 2, 2020 · Terminated
    7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
  • Z-0958-2021 · Dec 2, 2020 · Terminated
    154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
  • Z-0956-2021 · Dec 2, 2020 · Terminated
    60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
  • Z-0955-2021 · Dec 2, 2020 · Terminated
    60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
  • Z-0183-2021 · Jul 6, 2020 · Terminated
    6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave T-Connector, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case; UDI:(01)10887709051024(17)250101(30)50(10)
  • Z-0239-2021 · Jul 6, 2020 · Terminated
    APPX 0.75 ml, 6-Port NanoClave Manifold, Check Valve. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709070551(17)241201(30)50(10)4542066; (01)10887709070551(17)241201(30)50(1
  • Z-0247-2021 · Jul 6, 2020 · Terminated
    7 IN (18CM) APPX 0.26ml, SMALLBORE EXT SET, MicroClave, Clave T-Connector.1 unit per pouch, 50 pouches per case. UDI: (01)10887709018065(17)250101(30)50(10)4572133. The Intravasc
  • Z-0190-2021 · Jul 6, 2020 · Terminated
    9 IN(23 cm) Appx 0.89ml, Smallbore Quadfuse, 4 MicroClave Clear, NanoClave, 0.2 Micron Filter;1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053004(17)250101(30)50
  • Z-0245-2021 · Jul 6, 2020 · Terminated
    126 IN (320cm) APPX 23.5ml, 10 Drop Set 3 Clave, Remv Clave Stopcock, 2 Ext.1 unit per pouch, 25 pouches per case. UDI:(01)10887709062365(17)250201(30)25(10)4630752. The Intrav
  • Z-0222-2021 · Jul 6, 2020 · Terminated
    123 IN (312cm) APPX 16.1 ml, 10 DROP SET w/3 MicroClave, 3-Port NanoClave; UDI:(01)10887709085432(17)241201(30)25(10)4536733;(01)10887709085432(17)250101(30)25(10)4573641. The In
  • Z-0241-2021 · Jul 6, 2020 · Terminated
    6 IN (15cm) APPX 0.89ml, Smallbore Set 6-Port NanoClave Manifold.1 unit per pouch, 50 pouches per case. UDI:(01)10887709087719(17)250101(30)50(10)4558621;(01)10887709087719(17)2501
  • Z-0196-2021 · Jul 6, 2020 · Terminated
    7.5" (19 cm) Appx 0.59 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear (Yellow, Green, Purple Rings), NanoClave T-Connector, 4 Clamps, Rotating Luer. UDI:(01)10887709067049(17)2

…and 172 more.

Other Stopcock, i.V. Set devices

All Stopcock, i.V. Set devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page