ICU Medical B9535Recalls & reported problems

Icu Medical, Inc. · 104" (264 cm) Appx 15.7 mL, 10 Drop Primary Set w/BCV-Clave®, Clave®, 0.2 Micron Filter, Rotating Luer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887709014343), received between Mar 23, 2022 and Mar 24, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

187 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 12, 2026Class IIOpen, Classified

    Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.

    Z-3163-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Contamination with Chemical or Other Material2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THE EVENT INVOLVED A 104" (264 CM) APPX 15.7 ML, 10 DROP PRIMARY SET W/BCV-CLAVE®, CLAVE®, 0.2 MICRON FILTER, ROTATING LUER THAT THE CUSTOMER REPORTED PARTICULATE ON THE TUBING UPON OPENING AND THE EVENT WAS NOTICED PRIOR TO ATTACHING ON THE PATIENT. THE TUBING SET WAS CONNECTED TO A MINIBAG PLUS SYSTEM. THERE WAS NO PATIENT INVOLVEMENT AND NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13874446 ↗
  • MalfunctionMar 23, 2022Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    THE EVENT INVOLVED 104" (264 CM) APPX 15.7 ML, 10 DROP PRIMARY SET W/BCV-CLAVE®, CLAVE®, 0.2 MICRON FILTER, ROTATING LUER. IT WAS REPORTED THAT THERE WAS A CHUNK OF DARK BROWN MATERIAL WAS STICKING TO THE INSIDE OF THE INTRAVENOUS (IV) LINE (NEAR THE END OF THE LINE). THIS WAS NOTICED AFTER THE LINE WAS PREPARED WITH MEDICATION (AMPICILLIN). ONCE THE LIQUID MEDICATION FLOWED PAST THE BROWN CHUNKS, SOME OF THE PARTICLES BROKE AWAY AND IT WAS REPORTED THAT APPROXIMATELY A DOZEN WERE FLOATING IN THE LINE. THE MAIN CHUNK OF BROWN MATERIAL CONTINUED TO STAY ADHERED TO THE INSIDE OF THE LINE. THE IV LINE WAS SEALED IN THE PACKAGING WHEN IT WAS RECEIVED AND IT WAS OPENED ABOUT 6PM. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE IS NO REPORT OF HUMAN HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13864874 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.

  • Z-3157-2026 · May 12, 2026 · Open, Classified
    IV Non-Dedicated Gravity Administration Sets with drip chambers
  • Z-0154-2026 · Oct 2, 2025 · Open, Classified
    IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
  • Z-1721-2022 · Aug 1, 2022 · Open, Classified
    6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
  • Z-0956-2021 · Dec 2, 2020 · Terminated
    60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
  • Z-0957-2021 · Dec 2, 2020 · Terminated
    7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
  • Z-0955-2021 · Dec 2, 2020 · Terminated
    60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
  • Z-0958-2021 · Dec 2, 2020 · Terminated
    154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
  • Z-0217-2021 · Jul 6, 2020 · Terminated
    Appx 0.82 ml, 3 Gang 4-Way NanoClave Stopcock w/Baseplate, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709084435(17)250101(30)50(10)
  • Z-0185-2021 · Jul 6, 2020 · Terminated
    5 IN(13cm) APPX 0.87ml, EXT SET,NanoClave T-CONNECTOR; 1 unit per pouch, 50 pouches per case; UDI: (01)10887709052809(17)250101(30)50(10)4599980.
  • Z-0191-2021 · Jul 6, 2020 · Terminated
    5" (13 cm) Appx 0.39 ml, Smallbore Bifuse Ext Set w/NanoClave T-Connector, 2 Clamps, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053028(17)250101
  • Z-0227-2021 · Jul 6, 2020 · Terminated
    142 IN (361cm) APPX 18.8 ml, 15 DROP SET 4 MicroClave CLEAR, REMV 3-PORT NanoClave w/ BCV. 1 unit per pouch, 25 pouches per case. UDI:(01)10887709089393(17)250101(30)25(10)4573948
  • Z-0195-2021 · Jul 6, 2020 · Terminated
    109" (277 cm) Appx 8.6 ml, Transfer Set w/Check Valve, NanoClave T-Connector, Anti-Siphon Valve, 4 Clamps, Smallbore Trifuse Ext Set w/2 MicroClave Clear, 0.2 Micron Filter, Rotati
  • Z-0187-2021 · Jul 6, 2020 · Terminated
    7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)108877090510
  • Z-0242-2021 · Jul 6, 2020 · Terminated
    11 IN (28cm) APPX 1.5ml EXT SET NanoClave, 6-Port NanoClave Manifold . 1 unit per pouch, 50 pouches per case. UDI:(01)10887709098937(17)241201(30)50(10)4493669;(01)10887709098937(
  • Z-0214-2021 · Jul 6, 2020 · Terminated
    99 IN (251cm) APPX 12.6ml, 20 Drop, 2 Clave, 2 NanoClave 4-Way Stopcocks, Drop-In Ext, Clave; 1 unit per pouch, 25 pouches per case. UDI: (01)10887709066318(17)241201(30)25(10)4515

…and 172 more.