ICU Medical Z3054Recalls & reported problems
Icu Medical, Inc. · 94" (239cm) 15 Drop Primary Set w/BCV-Clave®, Clave®, 3 Gang 1o2® Manifold, Check Valve, Rotating Luer, 2 Ext · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887709029330), received between Jun 17, 2024 and Jun 17, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
188 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2024Leak/Splash; Leak/Splash
COMPLAINT RECEIVED REPORTING LEAKAGE ISSUES WITH USE OF Z3054, 94" PRIMARY DROP SET W/BCV-CLAVE, CLAVE, 3 GANG 1O2, LOT# UNKNOWN. IT WAS REPORTED THAT "PRODUCT SET IN PRE-OP, IN OR SETTING (3 GANG 1O2) MANIFOLDS HAVE LEAKED[?] IN ALL SITUATIONS THE PROCEDURE HAS BEEN COMPLETED AND THE TUBING TAKEN OFF AFTER SURGERY...". THERE WERE NO REPORTED ADVERSE PATIENT CONSEQUENCES AND OR OUTCOMES. DEVICE RETURN: ONE USED Z3054 DEVICE SET ATTACHED TO UNKNOWN EXT. SET AND A TOTAL OF SIX PACKAGED Z3054 DEVICE SETS FROM "POTENTIAL" LOT# WERE RECEIVED ENGINEERING TESTING AND ANALYSIS WAS PERFORMED. THE RETURNED USED Z3054 AND THE SIX UNUSED Z3054 DEVICE SETS WERE TESTED TO THE APPLICABLE PRODUCT SPECIFICATIONS. THE RESULTS RECORDED FOR THE USED Z3054 CONFIRMED LEAKAGE WAS REPLICATED ORIGINATING FROM THE DISTAL PORT ON THE SETS 3 GANG 1O2 MANIFOLD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19554700 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.
- Z-3157-2026 · May 12, 2026 · Open, ClassifiedIV Non-Dedicated Gravity Administration Sets with drip chambers
- Z-3163-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-0154-2026 · Oct 2, 2025 · Open, ClassifiedIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
- Z-1721-2022 · Aug 1, 2022 · Open, Classified6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
- Z-0955-2021 · Dec 2, 2020 · Terminated60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
- Z-0958-2021 · Dec 2, 2020 · Terminated154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
- Z-0956-2021 · Dec 2, 2020 · Terminated60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
- Z-0957-2021 · Dec 2, 2020 · Terminated7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
- Z-0231-2021 · Jul 6, 2020 · Terminated10 IN (25cm) APPX 1.5ml, Ext w/3-Port NanoClave Manifold, Check Valve. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709080796(17)241201(30)50(10)4536743; (01)1088770908
- Z-0195-2021 · Jul 6, 2020 · Terminated109" (277 cm) Appx 8.6 ml, Transfer Set w/Check Valve, NanoClave T-Connector, Anti-Siphon Valve, 4 Clamps, Smallbore Trifuse Ext Set w/2 MicroClave Clear, 0.2 Micron Filter, Rotati
- Z-0187-2021 · Jul 6, 2020 · Terminated7" (18cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave Clear, NanoClave T-Connector, Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)108877090510
- Z-0256-2021 · Jul 6, 2020 · Terminated149 IN (378cm) APPX 19.1 ml, 15 DROP SET, 3 Clave, REMV 3 GANG 4-WAY Clave STOPCOCK 2 EXT. 1 unit per pouch, 25 pouches per case. UDI:(01)10887709074696(17)250101(30)25(10)4711277;
- Z-0253-2021 · Jul 6, 2020 · Terminated60 IN(152cm) APPX 0.85ml, SMALLBORE, CLAVE STOPCOCK, MicroCLAVE CLEAR. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709040264(17)250201(30)50(10)4714442. The Intravascular
- Z-0211-2021 · Jul 6, 2020 · Terminated5.5" (14 cm) Appx 0.50 ml, Smallbore Trifuse Ext Set w/3 MicroClave" Clear, NanoClave" T-Connector, 3 Clamps, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case.UDI:(01)
- Z-0246-2021 · Jul 6, 2020 · Terminated40 (102cm) Appx 4.4ml Ext Set, 4-Way Clave Stopcock, Clave, 2 Clamps, Rotating Luer, 1 Ext. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709014531(17)250101(30)50
…and 173 more.
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.