ICU Medical A1003Recalls & reported problems
Icu Medical, Inc. · 6" (15 cm) Appx 0.24 ml, Smallbore Ext Set w/MicroClave® Clear, Clamp, NanoClave® T-Connector, Rotating Luer · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887709051034), received between Aug 9, 2023 and Mar 5, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
188 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT A 6" (15 CM) APPX 0.24 ML, SMALLBORE EXT SET W/MICROCLAVE® CLEAR, CLAMP, NANOCLAVE® T-CONNECTOR, ROTATING LUER WAS FOUND TO LEAK DURING PATIENT USE. IT WAS STATED IN THE REPORT THAT THE PATIENT'S WHITE LUMEN WAS LEAKING AS EVIDENCED BY WETNESS ON THE CENTRAL VENOUS LINE (CVL) VEST AS WELL AS SEEN WHEN FLUSHING. THE CAP AND TUBING CHANGE WAS COMPLETED IMMEDIATELY, AND THE WHITE LUMEN WAS CAPPED ALTHOUGH THE LUER LOCK WAS STILL NOT FULLY SECURE. WHITE LUMEN WAS NOT USED FOR THE REMAINDER OF THE SHIFT AND A COMPUTERIZED AXIAL TOMOGRAPHY( CAT) TEAM CONSULT WAS PLACED TO EVALUATE THE ISSUE IN THE MORNING. THE CAT TEAM REVIEWED AND INDICATED THE ISSUE WAS THE T CONNECTOR WAS SWIVELING AND NOT CONNECTING CORRECTLY TO THE LUMEN. THE T CONNECTOR DID NOT HAVE ENOUGH RIDGES TO TIGHTEN ONTO THE END OF THE LUMEN. THEIR TEAM NOTICED THE ISSUE SOME TIME AGO THAT THE T CONNECTORS WEREN¿T FITTING WELL. THERE WAS NO PATIENT HARM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21528887 ↗MalfunctionAug 9, 2023Material Separation; Material Separation
THE INCIDENT INVOLVED A 6" (15 CM) APPX 0.24 ML, SMALLBORE EXT SET W/MICROCLAVE® CLEAR, CLAMP, NANOCLAVE® T-CONNECTOR, ROTATING LUER. IT WAS REPORTED THAT THE CLAVE VALVE DISENGAGED FROM THE MECHANISM. THE EVENT OCCURRED WHEN CHANGING FLUIDS AND WAS NOTICED IMMEDIATELY UPON CONNECTING. IT WAS THEN REMOVED AND A SECOND PRODUCT WAS USED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17491221 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.
- Z-3157-2026 · May 12, 2026 · Open, ClassifiedIV Non-Dedicated Gravity Administration Sets with drip chambers
- Z-3163-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-0154-2026 · Oct 2, 2025 · Open, ClassifiedIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
- Z-1721-2022 · Aug 1, 2022 · Open, Classified6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
- Z-0955-2021 · Dec 2, 2020 · Terminated60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
- Z-0957-2021 · Dec 2, 2020 · Terminated7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
- Z-0958-2021 · Dec 2, 2020 · Terminated154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
- Z-0956-2021 · Dec 2, 2020 · Terminated60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
- Z-0230-2021 · Jul 6, 2020 · Terminated114" (290 cm) 60 Drop 150ml Burette Set (Clave, Shut Off, Filter) w/4 Clave Remv 3-Port NanoClave Manifold, 2 Check Valves, Rotating Luer, 2 Exts. 1 unit per pouch, 25 pouches per
- Z-0202-2021 · Jul 6, 2020 · Terminated8" (20 cm) Appx 0.53 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Filter, NanoClave T-Connector, Clamp, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(
- Z-0183-2021 · Jul 6, 2020 · Terminated6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave T-Connector, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case; UDI:(01)10887709051024(17)250101(30)50(10)
- Z-0241-2021 · Jul 6, 2020 · Terminated6 IN (15cm) APPX 0.89ml, Smallbore Set 6-Port NanoClave Manifold.1 unit per pouch, 50 pouches per case. UDI:(01)10887709087719(17)250101(30)50(10)4558621;(01)10887709087719(17)2501
- Z-0196-2021 · Jul 6, 2020 · Terminated7.5" (19 cm) Appx 0.59 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear (Yellow, Green, Purple Rings), NanoClave T-Connector, 4 Clamps, Rotating Luer. UDI:(01)10887709067049(17)2
- Z-0224-2021 · Jul 6, 2020 · Terminated3-Port NanoClave Manifold w/Check Valve, Rotating Luer. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709087153(17)250101(30)50(10)4556715; (01)10887709087153(17)250101(30)50
- Z-0259-2021 · Jul 6, 2020 · Terminated5.5" (14 cm) Appx 0.39 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, NanoClave T-Connector, 3 Clamps, Rotating Luer.1 unit per pouch, 50 multivac pouches per case. UDI: (01)1
…and 173 more.
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page