ICU Medical B1729Recalls & reported problems
Icu Medical, Inc. · 5.5" Ext Set w/0.2 Micron Filter, Luer Lock · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887709073951), received between Jul 23, 2020 and May 24, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
188 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Obstruction of Flow1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A 5.5" EXT SET W/0.2 MICRON FILTER, LUER LOCK WHERE IT WAS REPORTED THAT DURING THE FIRST TITRATION STEP, PATIENT STATED HE NOTICED HIS MICRON FILTER WAS LEAKING AND PILLOWCASE HE WAS RESTING HIS ARM ON WAS WET. INFUSION WAS IMMEDIATELY STOPPED. THERE WAS PATIENT INVOLVEMENT AND NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19390695 ↗MalfunctionOct 16, 2020Obstruction of Flow
FILTER ATTACHMENT TO IV TUBING OCCLUDED, FLUID WOULD NOT RUN DURING SIMPONI ARIA INFUSION. CHANGED PUMPS, CHECKED ALL PORTS AND EQUIPMENT. WHEN REQUIRED FILTER ATTACHMENT WAS REMOVED THE OCCLUSION WAS GONE. FILTER ATTACHMENT WAS REPLACED AND INFUSION WAS COMPLETED WITHOUT FURTHER COMPLICATION. FILTER ATTACHMENT PART REFERENCE NUMBER B1729, LOT 4799732, EXPIRATION 04/01/2025.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10689709 ↗MalfunctionJul 23, 2020Positioning Problem
THERE HAS BEEN THREE SEPARATE EVENTS IN THE NICU INVOLVING THE USE OF THE ICU MEDICAL B1729 LUER LOCK EXTENSION SET WITH 0.2 MICRON FILTER. THE FIRST EVENT ((B)(4)) NOTED THE MALFUNCTIONING OF A B1729 IV EXT SET CONNECTED TO AN UMBILICAL VENOUS CATHETER (UVC). WHEN THE MICRON FILTER WAS PLACED AT THE SAME LEVEL AS THE NEWBORN PATIENT, THE UVC BLED BACK INTO THE IV LINE AND TUBING. THE MICRON FILTER WORKED AS EXPECTED WHEN HELD IN THE "HOUSE UP" POSITION. THE IV FLOW RATE DURING THIS EVENT WAS 17.5 ML/HR. THE ISSUE WITH THE B1729 EXT SET STARTED WHEN THE FILTER HAD TO BE ADDED TO THE NO-PORT IV TUBING. THE 0.2 FILTER IS SHORTER THAN THE FILTERS USED IN THE PAST WHICH ARE 14 CM IN TOTAL LENGTH. THE SECOND EVENT ((B)(4)) INVOLVED BLOOD BACKING UP IN THE UVC LINE USING THE B1729 MICRON FILTER CONNECTED TO A BD ALARIS IV INFUSION SET. THE BD IV PUMP WAS USED TO INFUSE FLUIDS AT A RATE OF 7.3 DURING A TOTAL PARENTERAL NUTRITION (TPN) PROCEDURE. RN NOTED THE PRESENCE OF BLOOD IN THE LINE AND CALLED THE PHYSICIAN. HEALTH CARE PROVIDERS NOTED THAT FLUID STARTED TO FLOW NORMALLY ON THE IV SET WHEN THE FILTER WAS HELD IN A VERTICAL POSITION (UP HIGHER IN THE AIR). BLOOD STARTED BACKING UP IN
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10312078 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.
- Z-3163-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-3157-2026 · May 12, 2026 · Open, ClassifiedIV Non-Dedicated Gravity Administration Sets with drip chambers
- Z-0154-2026 · Oct 2, 2025 · Open, ClassifiedIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
- Z-1721-2022 · Aug 1, 2022 · Open, Classified6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
- Z-0958-2021 · Dec 2, 2020 · Terminated154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
- Z-0957-2021 · Dec 2, 2020 · Terminated7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
- Z-0956-2021 · Dec 2, 2020 · Terminated60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
- Z-0955-2021 · Dec 2, 2020 · Terminated60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
- Z-0254-2021 · Jul 6, 2020 · Terminated22" (56 cm) Appx 2.5 ml, Ext Set w/4-Way Clave Stopcock, Clamp, Luer Lock. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709051967(17)250101(30)50(10)4605591; (01)10
- Z-0253-2021 · Jul 6, 2020 · Terminated60 IN(152cm) APPX 0.85ml, SMALLBORE, CLAVE STOPCOCK, MicroCLAVE CLEAR. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709040264(17)250201(30)50(10)4714442. The Intravascular
- Z-0251-2021 · Jul 6, 2020 · Terminated60 IN (152cm) APPX 1.1ml, TRIFUSE, Clave STOPCOCK, 2 Clave, FILTER. 1 unit per pouch, 50 pouches per case. UDI:(01)10887709040103(17)250201(30)50(10)4726780. The Intravascular Ad
- Z-0226-2021 · Jul 6, 2020 · Terminated112 IN(284cm) APPX 14.6ml, 15 Drop Set 3 MicroClave, 3-Port NanoClave Manifold, BCV.1 unit per pouch, 25 pouches per case. UDI:(01)10887709088518(17)241101(30)25(10)4475764 (01)108
- Z-0258-2021 · Jul 6, 2020 · Terminated112" (284 cm) 60 Drop 150ml Burette Set (Clave, Shut-Off) w/2 MicroClave, Remv 3 Gang 4-Way Clave Stopcock, Check Valve, Rotating Luer, 1 Ext. 1 unit per pouch, 25 pouches per case
- Z-0202-2021 · Jul 6, 2020 · Terminated8" (20 cm) Appx 0.53 ml, Smallbore Ext Set w/MicroClave Clear, 0.2 Micron Filter, NanoClave T-Connector, Clamp, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(
- Z-0227-2021 · Jul 6, 2020 · Terminated142 IN (361cm) APPX 18.8 ml, 15 DROP SET 4 MicroClave CLEAR, REMV 3-PORT NanoClave w/ BCV. 1 unit per pouch, 25 pouches per case. UDI:(01)10887709089393(17)250101(30)25(10)4573948
…and 173 more.
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page