ICU Medical MC330310Recalls & reported problems
Icu Medical, Inc. · 14" (36 cm) Ext Set w/1.2 Micron Filter, MicroClave™ Clear, Clamp, Rotating Luer · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887709083189), received between Jan 10, 2025 and Jan 13, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
188 other recalls cover devices cleared under the same FDA submission (K964435); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Infusion or Flow Problem1
- Leak/Splash1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2025Leak/Splash; Leak/Splash
THE COMPLAINT/EVENT INVOLVED A 14" (36 CM) EXT SET W/1.2 MICRON FILTER, MICROCLAVE® CLEAR, CLAMP, ROTATING LUER. A LEAK WAS REPORTED FROM THE PORT ON THE LIPID TUBING DIRECTLY BELOW LIGHT BLUE ALCOHOL CAP. THE INFUSION LINE WAS STOPPED AND THE TUBING WAS DISCONNECTED. UPON DISCONNECTING THE INFUSION LINE AT THE BLUE CAP SITE PRIOR TO FLUSHING, THE NURSE NOTICED THE HUB ON THE CAP THAT CONNECTED TO THE PATIENT (NOT ON TUBING, ON DARK BLUE BIFURCATED EXTENSION) DID NOT LAY FLUSH. AS SUCH, IF ONE WERE TRYING TO SCRUB THE HUB THERE WAS AN INDENT AND IT WAS NOT FLAT. THERE WAS NO PATIENT HARM, BUT THE PATIENT DID NOT RECEIVE THE FULL DOSE OF PRESCRIBED LIPIDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21145918 ↗MalfunctionJan 13, 2025Infusion or Flow Problem; Infusion or Flow Problem
THE COMPLAINT/EVENT INVOLVED A 14" (36 CM) EXT SET W/1.2 MICRON FILTER, MICROCLAVE® CLEAR, CLAMP, ROTATING LUER. THE REPORTER STATED THAT SHE CLAMPED THE TOTAL PARENTERAL NUTRITION (TPN) AT THE QUADFUSE AND ATTEMPTED TO BEGIN INFUSING THE LIPIDS ON THE PUMP. THE LIPIDS NEVER INFUSED THROUGH THE CAP INTO THE QUADFUSE. UPON UNCLAMPING THE TPN, IT IMMEDIATELY BACK-FLOWED BACK UP INTO THE LIPID TUBING. THERE WAS NO PATIENT HARM, BUT THE PATIENT DID NOT RECEIVE THE FULL DOSES OF PRESCRIBED LIPIDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21145830 ↗MalfunctionJan 10, 2025Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
THE EVENT INVOLVED A 14" (36 CM) EXT SET W/1.2 MICRON FILTER, MICROCLAVE® CLEAR, CLAMP, ROTATING LUER WHERE A REPORT STATED THAT PARTS OF THE LIPID LINE TUBING BREAKING OR LEAKING. THE RN FOUND LIPID TUBING AT SITE NEAR FILTER DISCONNECTED AND BROKEN OFF. LIPIDS WERE CONTINUING TO RUN BUT NOT GOING TO PATIENT. WHEN FOUND, LIPIDS STOPPED, AND THE REST OF THE LINE WAS DISCONNECTED FROM THE PATIENT. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM, BUT ALL PATIENTS DID NOT RECEIVE THEIR FULL DOSES OF PRESCRIBED LIPIDS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21128693 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K964435. One submission often covers many models, so check the product and lot details in each record.
- Z-3157-2026 · May 12, 2026 · Open, ClassifiedIV Non-Dedicated Gravity Administration Sets with drip chambers
- Z-3163-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-0154-2026 · Oct 2, 2025 · Open, ClassifiedIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B
- Z-1721-2022 · Aug 1, 2022 · Open, Classified6.5" (17 cm) Appx 0.84 ml, Smallbore Bifuse Ext Set w/2 MicroClave Clear, 2 Check Valves, Luer Lock;
- Z-0958-2021 · Dec 2, 2020 · Terminated154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides ac
- Z-0957-2021 · Dec 2, 2020 · Terminated7 IN (18cm) APPX 0.43ml, SMALLBORE SET, MicroCLAVE, FILTER SetSource - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pro
- Z-0955-2021 · Dec 2, 2020 · Terminated60 IN (152cm) APPX 1.5 ml, PUR YELLOW SMALLBORE SET, FILTER REF B2209 - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which pr
- Z-0956-2021 · Dec 2, 2020 · Terminated60" (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave Clear, 1.2 Micron Filter, 2 Clamps, Luer Lock REF MC330174 - Product Usage: The intravascular administration set is a singl
- Z-0217-2021 · Jul 6, 2020 · TerminatedAppx 0.82 ml, 3 Gang 4-Way NanoClave Stopcock w/Baseplate, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709084435(17)250101(30)50(10)
- Z-0191-2021 · Jul 6, 2020 · Terminated5" (13 cm) Appx 0.39 ml, Smallbore Bifuse Ext Set w/NanoClave T-Connector, 2 Clamps, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053028(17)250101
- Z-0232-2021 · Jul 6, 2020 · Terminated58 IN (147cm) APPX 8.0 ml, 3 Port NanoClave Manifold NanoClave Stopcock, MicroClave Clear.1 unit per pouch, 25 pouches per case. UDI:(01)10887709095298(17)250101(30)25(10)4573460
- Z-0190-2021 · Jul 6, 2020 · Terminated9 IN(23 cm) Appx 0.89ml, Smallbore Quadfuse, 4 MicroClave Clear, NanoClave, 0.2 Micron Filter;1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053004(17)250101(30)50
- Z-0244-2021 · Jul 6, 2020 · Terminated128" (325 cm) Appx 16.6 ml, 15 Drop Primary Set w/2 MicroClave, Remv 2 Gang 4-Way Clave Stopcock, Rotating Luer, 2 Ext. 1 unit per pouch, 50 multivac pouches per case.UDI:(01)10887
- Z-0186-2021 · Jul 6, 2020 · Terminated7" (18 cm) Appx 0.29 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave, Clamp, Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709051048(17)250101(30)
- Z-0238-2021 · Jul 6, 2020 · Terminated11" (28 cm) Appx 1.7 ml, Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, NanoClave 4-Way Stopcock, Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)1
…and 173 more.
Other Stopcock, i.V. Set devices
- Connecta, Luer-Lok 394910BD Switzerland Sàrl · Stopcock, i.V. Set3 recalls8 reports
- Outlook® 490361B. Braun Medical Inc. · Stopcock, i.V. Set3 recalls0 reports
- Blood Administration Set 490530B. Braun Medical Inc. · Stopcock, i.V. Set2 recalls11 reports
- Medex MX9341LIcu Medical, Inc. · Stopcock, i.V. Set63 reports
- ICU Medical AC135Icu Medical, Inc. · Stopcock, i.V. Set1 recall10 reports
- IV Administration Set 490067B. Braun Medical Inc. · Stopcock, i.V. Set1 recall9 reports
- ICU Medical MC330174Icu Medical, Inc. · Stopcock, i.V. Set1 recall4 reports
- ICU Medical Z2587Icu Medical, Inc. · Stopcock, i.V. Set1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page