Skip to content

ICU Medical AC218Recalls & reported problems

Icu Medical, Inc. · 3 Gang 4-Way NanoClave™ Stopcock w/Baseplate, Rotating Luer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887709084438), received between Nov 20, 2017 and Nov 20, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Connection Problem” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K080077); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Connection Problem1

Patient problems

  • Chemical Exposure1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 20, 2017Connection Problem; Connection Problem

    SEPARATION/LEAKAGE REPORTED. CUSTOMER REPORTS THAT THE NANOCLAVE JUST POPS OFF WHEN THE SPIROS IS REMOVED. REPORT OF CHEMO SPILLAGE ON PHARMACIST'S HAND. DEVICE WAS RETURNED FOR EVALUATION. VISUAL INSPECTION REVEALED THAT PACKAGE CONTAINED THE DEVICE WITH A BROKEN PORT WHICH HAD EVIDENCE OF EXTERNAL FORCE APPLIED THROUGH THE PACKAGE.

    Patient problems: Chemical Exposure

    MAUDE report 7044792 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080077. One submission often covers many models, so check the product and lot details in each record.

  • Z-0213-2021 · Jul 6, 2020 · Terminated
    2 Gang 4-Way NanoClave Stopcock Manifold w/Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053172(17)241201(30)50(10)4493459;(01)10887709053172(17)24
  • Z-0250-2021 · Jul 6, 2020 · Terminated
    Clave Stopcock. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709039916(17)250201(30)50(10)4727496. The ICU Medical Clave Stopcock is a device used to administer o
  • Z-0220-2021 · Jul 6, 2020 · Terminated
    Appx 0.82 ml, 3 Gang 4-Way NanoClave Stopcock Manifold w/Rotating Luer. 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053189(17)241201(30)50(10)4537253;(01)108877
  • Z-0212-2021 · Jul 6, 2020 · Terminated
    NanoClave Stopcock w/NanoClave; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709062112(17)250101(30)50(10)4572842. The ICU Medical Clave Stopcock is a device used

Other Stopcock, i.V. Set devices

All Stopcock, i.V. Set devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page