Skip to content

Liko LiftPants™, Mod. 92 3592327Recalls & reported problems

Liko AB · Patient lifting system lift pants · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887761984028), received between Jan 14, 2025 and Jan 16, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIFTPANTS 92 PES MESH XL HAD AN ISSUE, THE LIFT PANTS HAVE BEEN INSPECTED DURING AND HAVE FAILED. THE INSPECTOR STATED THE STRAP WAS TESTEDAT APPROX. 100KG, AND THE STRAP CAME APART THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21166908 ↗
  • MalfunctionJan 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT LIFTPANTS 92 PES MESH XL, HAD AN ISSUE, DURING USE WITH A PATIENT THE STITCHING BECAME UNDONE AND CAUSED THE PATIENT TO DROP ABOUT 6 INCHES BACK ONTO THE CHAIR THEY WERE LIFTED FROM. NO INJURY WAS REPORTED. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21151908 ↗

Other Lift, patient, non-ac-powered devices

All Lift, patient, non-ac-powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page