This is a TruSystem operating table, a powered surgical table with orthopedic accessories. The 43 reports mostly describe the table moving on its own during surgery, software or controls stopping working, and one section falling off the table. Reports mention 1 injury and no deaths. There are no recalls for this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 43
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 43 adverse event reports that list this exact device (DI 00887761989719), received between Mar 8, 2021 and Aug 11, 2026 — classified as 1 injury, 42 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement18
- Operating System Becomes Nonfunctional10
- Key or Button Unresponsive/not Working4
- Device Fell3
- No Apparent Adverse Event3
- Appropriate Device Problem Term/Code Not Available2
- Electrical Shorting2
- Insufficient Device Problem Information2
Patient problems
- No Clinical Signs, Symptoms or Conditions42
- Liver Laceration(s)1
Recent adverse event reports
The 25 most recent of 43 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 11, 2026Unstable
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50212129 ↗MalfunctionMar 23, 2026Unintended Movement
ON (B)(6) 2026, A COMPANY REPRESENTATIVE/SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT OPERATING TABLE TRUSYSTEM 7000 (PRODUCT CODE 1723633, SERIAL NUMBER (B)(6)), THE TABLE IS GOING UP AND DOWN ON ITS OWN. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24667249 ↗MalfunctionDec 26, 2025Unintended Movement; Device Fell; Unintended Movement; Device Fell
BAXTER RECEIVED A REPORT STATING THAT DURING A SURGICAL PROCEDURE, A OPERATING TABLE TRUSYSTEM 7000 HAD LIFTED ONE FOOT OFF THE GROUND AT ONE END OF THE TABLE. THE CUSTOMER THEN REVERSED THE POSITION AND THE TABLE LOWERED RAPIDLY. THIS OCCURRED DURING PATIENT USE. THE PATIENT WAS UNINJURED, AND THE PROCEDURE WAS FINISHED WITHOUT INCIDENT. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23898920 ↗MalfunctionJun 25, 2025Device Fell; Device Fell
BAXTER RECEIVED A REPORT STATING THAT OR TABLE TRUSYSTEM 7000 U (DV) EAD SECTION FELL OFF OF THE TABLE DURING A SURGICAL PROCEDURE WITH A PATIENT ON THE TABLE. THE BED WAS LOCATED AT THE ACCOUNT AND NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22330871 ↗MalfunctionJun 24, 2025Unintended Movement; Unintended Movement
BAXTER RECEIVED A REPORT STATING THAT CUSTOMER WAS TRYING TO LOWER THE TABLE TRUSYSTEM 7000 U (DV) AND IT WAS NOT FUNCTIONAL. SUDDENLY, THE TABLE DROPPED 5-6". THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED DURING A CASE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22318255 ↗MalfunctionApr 2, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE TRUSYSTEM 7000 U (DV) HAD AN ISSUE, TABLE SHUT OFF IN THE MIDDLE OF A SURGERY WHILE IN THE TRENDELENBURG POSITION. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21754518 ↗MalfunctionMar 18, 2025Unintended Movement; Unintended Movement
CUSTOMER REPORTED TABLE MOVED DURING THE PROCEDURE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21631055 ↗MalfunctionMar 18, 2025Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
THE CUSTOMER REPORTED THEY WERE UNABLE TO LEVEL THE TABLE OR GO INTO TREND MODE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21630142 ↗MalfunctionMar 12, 2025Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
CUSTOMER REPORTED TABLE WILL NOT ARTICULATE WITH REMOTE OR KEYPAD DURING CASE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT#: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21584091 ↗MalfunctionMar 4, 2025Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
CUSTOMER REPORTED BED IS NOT WORKING. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #C-2330641
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21517602 ↗MalfunctionMar 4, 2025No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available; No Apparent Adverse Event; Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REPORTED THAT BRAKES DISENGAGE. THIS REPORT WAS FILLED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21514556 ↗MalfunctionFeb 21, 2025Unintended Movement; Unintended Movement
CUSTOMER REPORTED TABLE BREAKS UNLOCKED. NO HARM OR IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21432691 ↗MalfunctionFeb 20, 2025Unintended Movement; Unintended Movement
CUSTOMER REPORTED TABLE DROPPED WHEN ATTEMPTING TO LEVEL FROM TREND MODE. THERE WAS NO REPORT OF PATIENT INJURY OR PROCEDURAL DELAY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21419845 ↗MalfunctionFeb 18, 2025Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
CUSTOMER REPORTED TABLE LOST ALL FUNCTIONS. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21401145 ↗MalfunctionDec 2, 2024Unintended Movement; Unintended Movement
CUSTOMER ALLEGED THAT THE TS7000 SURGICAL TABLE WOBBLES. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20818375 ↗MalfunctionNov 27, 2024Unintended Movement; Unintended Movement
CUSTOMER REPORTED TABLE UNLOCKED DURING CASE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20794438 ↗MalfunctionNov 20, 2024Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
CUSTOMER REPORTED THAT THEY LOST CONTROLS ON REMOTE AND COLUMN KEYPAD. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20733617 ↗MalfunctionNov 4, 2024Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional
THE CUSTOMER ALLEGED THAT THEIR TS7000 DV SURGICAL TABLE HAS NO FUNCTION AND WILL NOT OPERATE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20597134 ↗MalfunctionOct 29, 2024Unintended Movement; Unintended Movement
CUSTOMER REPORTED THAT THE DV TABLE WAS LOCKED INTO PLACE AND THE TABLE TOP WAS TWITCHING ALTHOUGH THE REMOTE/COLUMN CONTROLS WEREN¿T BEING TOUCHED. THERE WAS NO REPORT OF PATIENT INJURY OR PROCEDURAL DELAY. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20560094 ↗MalfunctionOct 4, 2024Electrical Shorting; Operating System Becomes Nonfunctional; Electrical Shorting; Operating System Becomes Nonfunctional
CUSTOMER REPORTED NO FUNCTIONS FROM REMOTE OR DISPLAY. CUSTOMER FOUND FLUID COMING FROM THE HAND PENDANT CONNECTOR AND ALSO FROM THE INSIDE OF THE TABLE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20372148 ↗MalfunctionSep 23, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
THE CUSTOMER REPORTED THAT WHEN TRYING TO MOVE THE OR TABLE, NO MOVEMENT WAS SEEN AND ALL THAT COULD BE HEARD WAS A MOTOR SOUND. THERE WAS NO PATIENT/USER INJURY REPORTED.THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20279433 ↗MalfunctionAug 12, 2024Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT WHEN PATIENT WAS ON THE TABLE, THE CASE WAS DELAYED BECAUSE THE TABLE WOULD NOT MOVE. THE TRUSYSTEM 7000DV OPERATING TABLE DRIVE MALFUNCTIONED AND TABLE WAS LOCKED BUT SHOWING UNLOCKED. ADDITIONALLY, AN ERROR CODE 429 WAS REPORTED. THE TABLE WAS BEING USED ON BOTH BATTERY AND AC WITH NO CHANGE. THE CUSTOMER RESET THE TABLE WITH NO CHANGE AND THE TABLE WOULD NOT LOCK OR UNLOCK WITH THE KEYPAD OR REMOTE. NO PATIENT INJURY WAS REPORTED, AND THE PATIENT WAS TRANSFERRED TO A DIFFERENT TABLE. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19962858 ↗MalfunctionAug 8, 2024Electrical Shorting; Key or Button Unresponsive/not Working; Electrical Shorting; Key or Button Unresponsive/not Working
THE CUSTOMER REPORTED THAT DURING A SURGICAL PROCEDURE WITH A PATIENT ON THE TABLE, THEIR TS7000 DV SURGICAL TABLE WOULD NOT ALLOW THEM TO MOVE OUT OF TRENDELENBURG POSITION. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19940631 ↗InjuryJun 24, 2024Device Fell; Device Fell
IT WAS REPORTED THAT WHEN THE TRUSYSTEM 7000 OPERATING TABLE WAS PUT INTO TRENDELENBURG, THE TABLE ALLEGEDLY DROPPED MAKING A LOUD NOISE AND CAUSING THE CAMERA TO NICK THE PATIENT¿S LIVER. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: Liver Laceration(s)
MAUDE report 19600738 ↗MalfunctionJun 4, 2024Unintended Movement; Unintended Movement
THE CUSTOMER REPORTED THAT THEIR TS7000 DV SURGICAL TABLE RANDOMLY UNLOCKED DURING A PROCEDURE. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19458599 ↗
Other TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED devices
- TS 7000 1604788Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall4 reports
- TS 7500 1717023Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall1 report
- TS 7500 1704695Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854085Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854087Baxter Medical Systems GmbH + Co. KG · TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- ADVANCED CONTROL I-BASE, DC 5803Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered47 reports
- TABLE, HANA, 110 V 6875Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered2 deaths36 reports
- Levó Arm 7887-050Mizuho Orthopedic Systems, Inc. · Table, Surgical With Orthopedic Accessories, Ac-Powered12 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page